Lack of validated blood pressure devices for use in pregnancy available from Australian pharmacies
Kaylee Slater1, Hollie Speer2, Niamh Chapman1,3
1Sydney School of Health Sciences, Faculty of Medicine and Health, University of Sydney, Sydney, NSW, Australia.
Blood pressure monitoring is a critical aspect of prenatal care, as hypertension during pregnancy can lead to serious complications such as preeclampsia, eclampsia, and other hypertensive disorders. Automatic blood pressure devices are widely used for home monitoring due to their convenience and ease of use. The use of validated automated blood pressure monitors is recommended by the International Society of Hypertension for home blood pressure measurements, and automatic devices require accuracy validation among people who are pregnant before they are recommended for use in pregnancy. This study evaluated availability of such devices from 18 Australian pharmacies. Only four devices (4/54, 7%) were validated for pregnancy and were more expensive than devices validated for the general population (14/54, 26%) and non-validated devices (40/54, 74%). Additionally, limited labelling and information was available to assist consumers to make informed purchasing decisions about home blood pressure devices for use in pregnancy. Increased availability, clear labelling and consumer education could help ensure use of appropriate blood pressure devices in pregnancy. Automatic blood pressure devices require additional accuracy validation for use in pregnancy. We found only four devices (4/54, 7%) were validated for pregnancy, which were more expensive than non-validated devices. Increased education could help ensure use of appropriate blood pressure devices in pregnancy.
Blood pressure monitoring is a critical aspect of prenatal care, as hypertension during pregnancy can lead to serious complications such as preeclampsia, eclampsia, and other hypertensive disorders. Automatic blood pressure devices are widely used for home monitoring due to their convenience and ease of use. The use of validated automated blood pressure monitors is recommended by the International Society of Hypertension for home blood pressure measurements, and automatic devices require accuracy validation among people who are pregnant before they are recommended for use in pregnancy. This study evaluated availability of such devices from 18 Australian pharmacies. Only four devices (4/54, 7%) were validated for pregnancy and were more expensive than devices validated for the general population (14/54, 26%) and non-validated devices (40/54, 74%). Additionally, limited labelling and information was available to assist consumers to make informed purchasing decisions about home blood pressure devices for use in pregnancy. Increased availability, clear labelling and consumer education could help ensure use of appropriate blood pressure devices in pregnancy. Automatic blood pressure devices require additional accuracy validation for use in pregnancy. We found only four devices (4/54, 7%) were validated for pregnancy, which were more expensive than non-validated devices. Increased education could help ensure use of appropriate blood pressure devices in pregnancy.
More Related Videos
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure
Pre-Procedural Guidelines for Assessing Blood Pressure
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure: