A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Reducing Unnecessary Implantable Cardioverter-Defibrillator Therapy With ENHANCED Programming: Long-Term Outcomes of
Cheyenne S L Chiu1, Akke van Dijkhuizen1, Willem Gerrits1
1Department of Cardiology, Division Heart & Lungs, University Medical Center Utrecht, Utrecht, the Netherlands.
Background:
Advances in implantable cardioverter-defibrillator (ICD) programming strategies have achieved significant reductions in inappropriate shocks. However, further refinement is needed to minimize appropriate but unnecessary therapies. The ENHANCED-ICD study initially demonstrated the short-term safety and efficacy of programming a number of intervals to detect (NID) of 60/80 over a median follow-up of 1.3 years. A decade later, this study presents the long-term impacts of this programming strategy.
Objective:
To assess the long-term impact of programming NID 60/80 for ventricular tachycardia (VT)/ventricular fibrillation (VF) detection on adverse events related to shocks and arrhythmias, as well as on ICD therapies-both delivered and avoided.
Methods:
A retrospective analysis was conducted on 60 patients from the ENHANCED-ICD study, a prospective, single-center trial. The median age was 60 years, 78% were men, and 53% had a primary prevention ICD indication. A prolonged detection interval of NID 60/80 was programmed for VT/VF detection. The cycle lengths for VT/fast VT/VF were set at 360/330/240 ms, respectively.
Results:
After a median follow-up of 9.4 years, Enhanced programming prevented unnecessary ICD therapies in 16.7% of patients and reduced the overall therapy rate by 25.9%. A total of 26.7% of patients received ICD therapy, with appropriate therapy delivered in 23.3% and inappropriate therapy occurring in 3.3% of patients. No arrhythmic deaths were observed, while syncope was reported in 10.0% of patients (1.63 per 100 patient-years).
Conclusion:
Prolonging the ICD detection interval to an NID of 60/80 successfully prevented appropriate but unnecessary therapy, while maintaining safety during long-term follow-up.
Trial Registration:
The ENHANCED-ICD study registered on http://ClinicalTrials.gov under study ID NCT01715116.
More Related Videos
10:17Real-Time Cardiac Mapping with a Noninvasive Imageless Electrocardiographic Imaging System
Published on: April 11, 2025
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
Related Concept Videos
Cardiomyopathy V: Interprofessional Care
Cardiomyopathy II: Dilated Cardiomyopathy
Heart Failure VI: Adjunct Therapies
Cardiomyopathy III: Hypertrophic Cardiomyopathy
Heart Failure V: Medical Management
Cardiomyopathy VII: Pre and Post Operative Nursing Management