Validation of blood pressure measurement using a smartwatch in patients with acute ischemic stroke
Yu-Hsuan Liu1, Li-Ching Chen2, Yi-Shian Guo3
1Department of Neurology, E-DA Hospital, Kaohsiung, Taiwan.
Abstract:
We aimed to validate the accuracy of blood pressure (BP) measurement using a smartwatch in patients with acute ischemic stroke. We compared 140 pairs of BP (n = 35) measurements acquired by a smartwatch (ASUS VivoWatch SP) with those measured by a sphygmomanometer (reference device). Differences between the smartwatch BP and reference BP measurements were compared. The validation procedure and criterion followed the consensus of the American National Standards Institute, Inc/Association for the Advancement of Medical Instrumentation/International Organization for Standardization (ANSI/AAMI/ISO) 81060-2:2018 and extended the standard to the specificities of cuffless devices in acute ischemic stroke population. The mean and standard deviation of the differences measured by a smartwatch and reference device were 1.8 ± 5.7 mm Hg in systolic BP and 0.7 ± 3.6 mm Hg in diastolic BP according to criterion 1. The mean and standard deviation of the differences measured by a smartwatch and reference device were 1.8 ± 5.6 mm Hg in systolic BP and 0.7 ± 3.6 mm Hg in diastolic BP according to criterion 2. The results both met the standards of the 2 criterion. The validation result did not differ between the paralytic and non-paralytic arms. The smartwatch with photoplethysmography sensors can provide accurate and reliable measurement of BP in acute ischemic stroke patients.
More Related Videos
05:51Assessing the Accuracy of Fitness Smartwatch Data for Cardiovascular and Physical Activity Monitoring: A Validation Study in Digital Health
Published on: February 21, 2025
09:00Real-Time Monitoring and Modulation of Blood Pressure in a Rabbit Model of Ischemic Stroke
Published on: February 10, 2023
Related Concept Videos
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Errors occurring during blood pressure monitoring
Several factors...
Pre-Procedural Guidelines for Assessing Blood Pressure
Measurement of Blood Pressure
Assessment of blood pressure in brachial artery(two-step method)