Related Experiment Video
Updated: Sep 10, 2025

Construction of a Wireless-Enabled Endoscopically Implantable Sensor for pH Monitoring with Zero-Bias Schottky Diode-based Receiver
Published on: August 27, 2021
Compact wideband implantable antenna for wireless capsule endoscopy application in the 2.45 GHz ISM band
1School of Electronics Engineering, Vellore Institute of Technology, Chennai, Tamil Nadu, India.
Abstract:
This paper presents a compact wideband implantable antenna developed for wireless capsule endoscopy (WCE) functioning in the 2.4-2.48 GHz ISM band. The antenna's miniaturized design is achieved through the integration of E - and L - shaped slots in the circular radiator, a shorting pin and defective ground structure with an inclined T - shaped slot. The final structure has compact dimension of [Formula: see text] and the Rogers RO3010 ([Formula: see text]) material serves as both substrate and superstrate. Performance was assessed in homogeneous muscle and heterogeneous human phantoms, and experimentally validated by implanting the device in minced pork. At 2.45 GHz the antenna delivered - 20.8 dBi gain, and a 44.02% impedance bandwidth. The specific absorption rate (SAR) values are recorded as 216.8 W/kg (1 - g) and 30.2 W/kg (10 - g). The study demonstrates that the antenna reliably supports wireless links beyond 10 m, maintaining a 10 dB margin at 2.45 GHz.