Summary
Radiation exposure during endoscopic lumbar discectomy is low for both interlaminar (IELD) and transforaminal (TELD) techniques. Both methods demonstrate comparable radiation doses, with TELD requiring longer fluoroscopy times.
Area of Science:
- Minimally Invasive Spine Surgery
- Medical Imaging
- Radiation Safety
Background:
- Spinal endoscopic procedures utilize fluoroscopy for instrumentation guidance.
- Fluoroscopy exposes medical staff and patients to ionizing radiation.
- Evaluating radiation doses in endoscopic lumbar discectomy is crucial for safety.
Purpose of the Study:
- To assess and compare radiation doses during interlaminar endoscopic lumbar discectomy (IELD) and transforaminal endoscopic lumbar discectomy (TELD).
- To analyze patient-specific and radiation-related variables for both procedures.
Main Methods:
- A retrospective analysis of 54 patients undergoing IELD (n=18) or TELD (n=36) for lumbar disc herniations.
- Data collected included patient demographics and radiation metrics: fluoroscopy time (FT), tube voltage (kVp), kerma-area product (KAP), and cumulative air-kerma (CAK).
Main Results:
- Significant differences observed in age, procedure duration, cine acquisitions, FT, KAP, and CAK between IELD and TELD groups.
- Mean KAP was 8.18 dGycm² for IELD vs. 29.09 dGycm² for TELD.
- Mean FT was 18.72 s for IELD vs. 54.94 s for TELD; skin doses remained below safe thresholds.
Conclusions:
- Intra-operative fluoroscopy during endoscopic discectomy yields low radiation doses.
- Radiation dose data were comparable between IELD and TELD techniques.
- TELD procedures necessitate longer fluoroscopy times and are more demanding.