Improved efficiency of the new Mirus™ system challenged by ventilatory settings-a bench study
Frederic Albrecht1, Claudia Schirra2, Kathrin Scheffler2
1Department of Anesthesiology, Intensive Care and Pain Therapy, Saarland University Medical Center and Saarland University Faculty of Medicine, Homburg, Germany. Frederic.Albrecht@uks.eu.
None:
The Mirus™ system (Technologie Institut Medizin GmbH, Germany) enables target-controlled delivery of volatile anaesthetics with open system ventilators. Its interface, positioned between Y-piece and the patient, injects anaesthetic vapor during inspiration, while adsorbing anaesthetic and resupplying it during the next inspiration. A newly introduced interface (Lisa-44) was evaluated under body temperature pressure saturated and normocapnic conditions using a test lung. Volume-controlled ventilation was applied with a 500 mL tidal volume, 10 bpm respiratory rate, and inspiratory flow (IF) times of 2.5 s and 0.5 s. Isoflurane (sevoflurane) target concentrations were stepwise increased to 1.5 MAC and then decreased. An external gas monitor recorded real-time anaesthetic concentrations. End-tidal concentrations (EC) and area under the curve (AUC) were analyzed for each breath. Accuracy and precision were assessed using Bland-Altman plots. Consumption was calculated and compared to historical controls. Mirus EC measurements correlated well with external gas monitor readings. ECs fluctuated around targets, with higher targets requiring more frequent injections and larger AUCs. Anaesthetic consumption was lower than historical controls and comparable to the Sedaconda ACD. At 0.3-0.5 MAC, hourly isoflurane consumption ranged from 1.0 to 1.7 mL (sevoflurane: 3.8-6.1 mL). However, very short IF times significantly increased consumption, reaching 43 mL sevoflurane at an EC of 2.6 vol%, with prolonged injections extending into the expiration phase. The new system demonstrates high accuracy, precision, and improved efficiency, suggesting reduced anaesthetic consumption in clinical use. However, high concentrations combined with very short IF times substantially increase consumption, indicating potential limitations.
Related Concept Videos
Mechanical Ventilation I: Indication and Settings
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...