Impact of Real-Time Feedback Devices on Paramedic Bag-Valve-Mask Ventilation Performance
Julian Lasik1, Tomasz Kłosiewicz1, Mateusz Puślecki1
1Chair of Emergency Medicine, Department of Medical Rescue, Poznan University of Medical Sciences, Poznan, Poland.
Objectives:
Ventilation using a self-expanding bag and face mask poses important challenges, even for experienced healthcare professionals. Several devices have been developed and introduced to the market with the aim of improving the quality of bag-valve-mask (BVM) ventilation. The objective of this study was to evaluate whether the use of a real-time ventilation feedback device (VFD) influences ventilation quality.
Methods:
This study recruited one hundred paramedics to perform manual ventilation on a high-fidelity manikin using a BVM, first without and subsequently with the VFD. Participants received a brief instruction on the use of VFDs. Each ventilation session lasted four minutes and was performed twice under both conditions. Ventilation parameters, including tidal volume and ventilation rate, were recorded using the SimMan manikin and LLEAP software.
Results:
Significant differences in ventilation parameters were observed between the two conditions. Ventilation without the VFD was associated with a higher mean tidal volume (390.1 mL vs. 373.9 mL, p < 0.001) and a higher mean ventilation rate (8.9 breaths per minute [bpm] vs. 8.1 bpm, p < 0.001). Minute ventilation was also significantly greater without the VFD (3504.0 mL vs. 3118.8 mL, p < 0.001). Despite these differences, the use of the VFD led to a slight improvement in compliance with European Resuscitation Council (ERC) ventilation guidelines, increasing the rate of correct ventilations from 1% to 3%.
Conclusions:
While the use of the VFD was associated with reductions in tidal volume and ventilation rate, it resulted in a modest improvement in adherence to ERC ventilation guidelines. The clinical impact of this device remains unclear. Future studies can focus on novel/innovative ways to enhance compliance with established ventilatory parameters.
Related Concept Videos
Cardiopulmonary Resuscitation II: ACLS Airway Management
Mechanical Ventilation I: Indication and Settings
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...