Dysfunctional Implantable Loop Recorder Post-Electrical Cardioversion: A Report of Two Cases
Wael Zaher1, Lorenzo Marcon1, Antonio Sorgente1
1Department of Cardiology Centre Hospitalier EpiCURA Hornu Boussu Belgium.
Implantable loop recorders (ILRs) may malfunction after electrical cardioversion. Careful anterior-posterior patch placement is crucial to prevent device dysfunction and ensure patient safety during the procedure.
Area of Science:
- Cardiology
- Biomedical Engineering
- Electrophysiology
Background:
- Implantable loop recorders (ILRs) are used for long-term cardiac rhythm monitoring.
- Electrical cardioversion is a common procedure to restore normal heart rhythm.
Purpose of the Study:
- To report cases of implantable loop recorder (ILR) dysfunction after electrical cardioversion.
- To investigate the potential impact of electrical cardioversion on ILR function.
Main Methods:
- Case report of two patients experiencing ILR dysfunction post-cardioversion.
- Analysis of device function in relation to electrical cardioversion procedure and lead placement.
Main Results:
- Two instances of implantable loop recorder (ILR) dysfunction were observed immediately following electrical cardioversion.
- Dysfunction was likely associated with the anterior-posterior placement of cardioversion patches.
Conclusions:
- Electrical cardioversion, particularly with anterior-posterior patch positioning, may pose a risk to implantable loop recorder (ILR) function.
- Careful consideration of patch placement is recommended to mitigate potential interference with cardiovascular implantable electronic devices like ILRs.
More Related Videos
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
08:36Surgical Implant Procedure and Wiring Configuration for Continuous Long-Term EEG/ECG Monitoring in Rabbits
Published on: January 24, 2025
Related Concept Videos
Dysrhythmias V: Evaluating Dysrhythmias
Cardiomyopathy V: Interprofessional Care
Dysrhythmias VI: Management of Dysrhythmias
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Cardiac Catheterization IV: Nursing Management