A novel device for temporary cardiac stimulation using permanent pacemaker generators and active-fixation leads: the
Raimundo Vicente-Miralles1,2, Nuria Rivas-Gándara3,4, Pablo Jordán Marchite3,4
1Intensive Care Unit, Dr. Balmis General University Hospital, Alicante, Spain.
Aims:
This study aims to determine the safety and effectiveness of the KronoSafe® permanent pacemaker adapter for temporary cardiac pacing (TCP) with active-fixation leads (TCPAFLs).
Methods And Results:
This was a multicentre, prospective, descriptive clinical investigation involving a medical device (ClinicalTrials.gov identifier: NCT05351658). Between January 2023 and December 2024, all consecutive patients who underwent implantation of a temporary active-fixation pacemaker were included. The devices were secured using the KronoSafe® adapter throughout the period of use. Complications during TCPAFL were recorded, and R-wave sensing, lead impedance, and pacing threshold were determined every 48 h. Thirty patients (10 with atrioventricular block, nine TAVI, five alcohol septal ablations, two slow atrial fibrillations, two bradycardia-tachycardia syndrome, and two pacemaker infections) who required TCP were included. Two centres participated in the study. The mean duration of TCP was 7.8 days (maximum 22 days), with 79.2% of the time spent in the cardiology ward. One complication (3%) was recorded, due to accidental traction of the system associated with an episode of agitation.
Conclusion:
Temporary cardiac pacing with active-fixation leads using the KronoSafe® adapter is safe and effective, and its use can be extended outside the intensive care unit.