Safety and Performance of INVICTA DF4 Ventricular Defibrillation Lead: Results From APOLLO Prospective Multicenter
Klaus K Witte1,2, Juan Gabriel Martínez3, Mário M Oliveira4
1Medizinische Klinik I Kardiologie Pneumologie Angiologie Und Internistische Intensivmedizin, Universitatsklinikum Aachen, Aachen, North Rhine-Westphalia, Germany.
The INVICTA DF4 lead shows high success in implanting and maintaining performance for defibrillator therapy. This quadripolar lead offers effective arrhythmia termination with low complication rates over 24 months.
Area of Science:
- Cardiology
- Medical Devices
- Electrophysiology
Background:
- Implantable cardioverter-defibrillator (ICD) therapy requires reliable right ventricular leads for crucial functions.
- The INVICTA DF4 lead is a quadripolar, active-fixation design engineered for improved implant workflow, safety, and long-term performance.
Purpose of the Study:
- To evaluate the safety and performance of the INVICTA DF4 lead in de novo ICD or CRT-D implantations.
- To assess long-term electrical variables, arrhythmia termination efficacy, and lead-related complications over 24 months.
Main Methods:
- The APOLLO study was a prospective, multicenter, single-arm observational trial involving 446 adult patients across Europe.
- Patients received either a 58 or 68 cm INVICTA DF4 lead for ICD or CRT-D implantation.
- Co-primary endpoints assessed safety and performance at 3 months; secondary endpoints evaluated long-term outcomes up to 24 months.
Main Results:
- Successful implantation was achieved in 100% of 446 patients, with leads placed in the apex (57.2%) or RV septum (42.8%).
- Electrical parameters remained stable post-discharge through 24 months, demonstrating durable performance.
- The lead achieved a 97.7% success rate in terminating arrhythmic episodes and a 96.4% overall lead-related complication-free rate.
Conclusions:
- The INVICTA DF4 ventricular lead demonstrated high implant success, sustained electrical performance, and effective arrhythmia management in a large European cohort.
- Low complication rates were observed through 24 months, with a trend towards fewer complications in septal lead positioning compared to apical placement.