Impact of Ultrahigh Sensitivity and Continuous Bed Motion on Performance Characteristics of Biograph Vision Quadra
Mostafa Roya1, Charalampos Tsoumpas2,3, Johannes H van Snick1
1Department of Nuclear Medicine and Molecular Imaging, University Medical Center Groningen, University of Groningen, Groningen, The Netherlands.
Abstract:
Continuous bed motion (CBM) for long-axial-field-of-view PET/CT enables swift total-body examination of patients. However, the sensitivity profile along the axial field of view (AFOV) varies significantly, and its effect on image quality when combined with CBM remains unexplored. This study assesses the effects of ultrahigh sensitivity (UHS) and CBM on recovery coefficients (RCs) and spatial resolution (SR). Methods: Phantom measurements, performed in static bed and CBM acquisition with different bed speeds (3.5, 7, 14, and 30 mm/s), were reconstructed for both high-sensitivity (HS) and UHS modes, and their quality was assessed using the RC and full width at half maximum of the point-spread function. In addition, 7 clinically referred oncologic patients underwent a combined CBM and UHS [18F]FDG PET/CT scan. Metrics derived from lesions and healthy tissues were compared across acquisition modes. Results: Mean RC was lower in both UHS (<4.2%) and CBM (<5.9%) than with HS and static acquisitions, respectively. SR in UHS was slightly deteriorated in the central plane of the AFOV (full width at half maximum difference range, 0.06-0.21 mm) but remained comparable toward the edge (full width at half maximum difference range, 0.03-0.07 mm) to SR in HS. In addition, for CBM, SR differences were negligible (full width at half maximum difference, <0.13 mm). SUVs in healthy tissues (r 2 > 0.94) and lesions (r 2 > 0.99) showed excellent correlation. The coefficient of variation was significantly different for the liver. Conclusion: Elaborate protocols using UHS and CBM could be a step toward swift total-body PET care with only a marginal increase in noise. CBM allows a larger part of the body to be examined and helps to mitigate substantial sensitivity differences along the AFOV, albeit at the cost of slightly lower RC.
More Related Videos
06:53Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
Published on: July 23, 2020
08:36Multi-Tracer Studies of Brain Oxygen and Glucose Metabolism Using a Time-of-Flight Positron Emission Tomography-Computed Tomography Scanner
Published on: June 7, 2024
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...