Related Experiment Video
Updated: May 3, 2026

Oral Biofilm Formation on Different Materials for Dental Implants
Published on: June 24, 2018
Comparative evaluation of retentive capacity of three different attachment systems for implant retained overdentures:
Radha Chauhan1, Narendra Padiyar1, Pragati Kaurani1
1Department of Prosthodontics and Crown & Bridge, Mahatma Gandhi Dental College and Hospital, Jaipur 302022, Rājasthān, India.
Background:
The primary issue in managing edentulous patients is the severely resorbed mandibular ridge, particularly in older individuals with diminished adaptive capacities. This compromised situation leads to the fabrication of inadequate dentures that lack retention and stability, potentially causing psychosocial issues.
Aim:
To determine the difference in retentive capacity between three attachment systems in implant-retained overdentures.
Methods:
Three edentulous mandibular models were fabricated using heat-cured polymethacrylate resin, with two implant replicas placed in the intra-foraminal region of each model. 30 acrylic resin mandibular overdentures were fabricated with provisions for three different overdenture attachment systems: A prefabricated ball/O-ring attachment, a locator attachment system, and an equator attachment system. Each model was subjected to 15000 pulls using a universal testing machine to remove the overdenture from the acrylic model and the force data were recorded.
Results:
The ball/O-ring attachment system demonstrated superior retentive capacity for 15 years, while the locator and equator attachment systems maintained excellent retentive capacity for 5 years.
Conclusion:
The ball/O-ring attachment system outperformed better than the other two attachment systems regarding retentive capacity. The locator and equator attachment systems presented sufficient retentive abilities until 15000 cycles. After 7500 cycles, significant differences in retentive force between the systems evolved.
More Related Videos
07:32Author Spotlight: 3D Movement Assessment of Maxillary Posterior Teeth in Clear Aligner Treatment
Published on: February 23, 2024
07:42Quasistatic Mechanical Testing for Computer-Aided Design and Manufacturing Occlusal Veneers Cemented to Milled Dentin Analog Material
Published on: December 20, 2024
Related Concept Videos
Insensitive Nuclei Enhanced by Polarization Transfer (INEPT)
Equivalence: In Vitro and In Vivo Bioequivalence