Photon-counting Detector CT Enables Pediatric Low-dose Chest Imaging With Further Reduction of Radiation Exposure
Background:
Photon-counting detector computed tomography (PCD CT) offers higher dose efficiency than conventional energy-integrating detector CT (EID CT), which is particularly beneficial for children. Broad evidence is missing whether frequently acquired pediatric low-dose lung imaging can be further improved using PCD CT.
Objective:
To compare radiation exposure, quantitative and qualitative image quality of pediatric low-dose chest PCD CT versus EID CT examinations.
Methods:
Unenhanced low-dose chest PCD CT and EID CT examinations acquired for clinical indications were retrospectively compared. Cohorts were matched by water-equivalent diameter (D w ) and age (n=44 each; median age 6.3 y PCD CT vs. 7.4 y EID CT). Radiation exposure was analyzed by volume CT dose index (CTDI vol ), dose length product (DLP), and size-specific dose estimate (SSDE). Quantitative image quality assessment featured the placement of regions of interest (ROIs) in the lung, heart, and liver for the extraction of mean attenuation, noise, signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), and figures of merit (FOMs). Qualitative image quality was evaluated by 3 readers using Likert scales and additional direct comparisons in a blinded manner.
Results:
Weight, height, and body mass index (BMI) were not significantly different between the 2 cohorts ( P >0.05). PCD CT examinations showed lower median CTDI vol (0.27 vs. 0.39 mGy, P <0.0001), DLP (6.71 vs. 8.75 mGy*cm, P <0.0001), and SSDE (0.55 vs. 0.83 mGy, P <0.0001) compared with EID CT. Mean attenuation [-797.76 vs. -772.50 Hounsfield units (HU), P =0.51], noise (17.82 vs. 17.69 HU, P =0.73), SNR (-46.10 vs. -45.40, P =0.63), and CNR (39.26 vs. 39.76, P =0.68) of lung parenchyma were not significantly different; respective dose efficiency expressed by FOM was higher in PCD CT compared with EID CT (mean 8030 vs. 5482 mGy -1 , P <0.0001). Qualitative rating showed equal and overall excellent scores for both cohorts.
Conclusions:
PCD CT enables pediatric low-dose chest imaging with lower radiation exposure at similar image quality compared with EID CT.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET