Parallel Cable-Driven Prostate Puncture Robot: Design and Performance Evaluation
Yupeng Zou1, Tianyu Lan1, Yanxing Zhang1
1College of Mechanical and Electronic Engineering, China University of Petroleum (East China), Qingdao, China.
Summary
A novel parallel cable-driven (PCD) robot simplifies prostate puncture surgery within MRI's magnetic field. This robotic system demonstrates high accuracy and stiffness, overcoming workspace limitations for improved surgical procedures.
Area of Science:
- Medical Robotics
- Surgical Technology
- Biomedical Engineering
Background:
- MRI environments pose significant challenges for robotic surgical system design due to limited workspace and strong magnetic fields.
- Simplifying robot structure is essential for effective prostate puncture procedures within MRI scanners.
Purpose of the Study:
- To propose and evaluate a novel parallel cable-driven (PCD) robot for prostate puncture surgery.
- To address the design challenges posed by MRI's magnetic field and limited workspace.
Main Methods:
- Development of a parallel cable-driven (PCD) prostate puncture robot with preliminary material design.
- Establishment of kinematics, statics, and stiffness models, with verification of the wrench-feasible workspace.
- Utilizing multibody simulation and motion accuracy experiments (planar and spatial) for performance analysis.
Main Results:
- The robot achieved high motion accuracy, with horizontal position errors of 0.35 mm and 0.4 mm during phantom puncture experiments.
- Stiffness performance was evaluated, demonstrating the robot's capability for precise surgical tasks.
- The parallel cable-driven design was shown to be compatible with strong magnetic fields.
Conclusions:
- The proposed non-parallel cable-driven design simplifies robotic structure for MRI-compatible prostate puncture surgery.
- The robot demonstrates high motion accuracy and superior stiffness, crucial for surgical precision.
- This innovative design overcomes MRI environmental challenges, enhancing potential for robotic-assisted prostate interventions.