Quality evaluation of synthetic CT using a gamma analysis for prostate cancer MR-only radiotherapy
Viktor Babka1, Lubos Rehounek2, Iva Jakubcova3
1Department of Oncology and Radiotherapy, University Hospital Hradec Kralove, Hradec Kralove, Czech Republic.
Background:
This study evaluates the quality of synthetic computed tomography (sCT) images for MR-only radiotherapy in prostate cancer using gamma analysis. A software tool, MRGamma, was developed to address challenges like the absence of electron density maps and registration uncertainties between magnetic resonance imaging (MRI) and planning CT (pCT).
Materials And Methods:
Aplication developed in MATLAB assesses Hounsfield units (HU) discrepancies between CT and sCT images via 2D and 3D gamma analysis (GA). GA computes a gamma index (γ) for each voxel based on HU differences and distance to agreement (DTA). The study analyzed prostate cancer patients using the MRCAT algorithm to generate sCT images. Preprocessing included registration, resampling, and cropping to standardize image dimensions. GA was performed with thresholds of 100 HU and 70 HU for HU differences and 3 mm for DTA to evaluate image quality.
Results:
Aplication demonstrated high agreement between sCT and CT images. For 3D GA (100 HU, 3 mm), 95.5 ± 1.8% of voxels passed the threshold, with slightly lower rates for 70 HU. Discrepancies were most pronounced at tissue interfaces and air cavities, where HU variations were more significant. Mean γ values for 3D GA were 0.19 ± 0.04 (milder parameters) and 0.23 ± 0.05 (stricter parameters), showing consistent trends in 2D GA. Maximum γ values confirmed good overall agreement.
Conclusions:
Aplication effectively evaluates sCT quality, supporting the feasibility of MR-only radiotherapy. By providing detailed HU comparisons, the tool enhances MR-based treatment planning, reducing costs and ionizing radiation exposure. Its implementation may improve workflow efficiency and patient safety in clinical practice.
More Related Videos
06:54MR Molecular Imaging of Prostate Cancer with a Small Molecular CLT1 Peptide Targeted Contrast Agent
Published on: September 3, 2013
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies III: Computed Tomography