Elucidating the Interplay Between Resin Media, Feed Constituents, and Process Conditions in AAV Affinity Column
Harshal Soni1,2, Sergey Pryshchep2, Thomas F Johnson3
1Department of Chemical and Biological Engineering, Rensselaer Polytechnic Institute, Troy, New York, USA.
Biotechnology and Bioengineering
|September 10, 2025
Summary
Confocal microscopy reveals differences in adeno-associated virus (AAV) resin loading and fouling. PSDVB-based resins show initial full access but suffer fouling upon reuse, unlike agarose-based resins.
Area of Science:
- Biotechnology
- Chromatography
- Microscopy
Background:
- Adeno-associated virus (AAV) purification relies on affinity chromatography.
- Understanding resin behavior during loading, elution, and cleaning is crucial for process optimization.
- Resin reuse impacts efficiency and purity in biopharmaceutical manufacturing.
Purpose of the Study:
- To investigate adeno-associated virus (AAV) loading and fouling in affinity resins using confocal microscopy.
- To compare the performance and reusability of PSDVB-based POROS CaptureSelect (PCS) and agarose-based AVIPure AAV9 resins.
- To elucidate the impact of resin characteristics and process conditions on column lifetime.
Main Methods:
- Confocal microscopy was used to examine AAV adsorption profiles in resin beads after various process steps.
- Commercial AAV affinity resins (PCS and AVIPure) were tested with different feed mixtures.
- High-resolution X-ray CT scans were employed to analyze resin pore structure.
Main Results:
- PCS resins allowed full AAV access initially, while AVIPure resins showed surface-limited binding.
- PCS resin fouling occurred upon reuse due to AAV and impurity carryover, leading to pore occlusion.
- AVIPure resins demonstrated effective AAV elution and removal during cleaning-in-place (CIP).
Conclusions:
- Resin pore size (PCS large pores) influences initial AAV transport.
- Fouling in PCS resins upon reuse significantly impedes solute transport in subsequent cycles.
- AVIPure resins exhibit better reusability and resistance to fouling compared to PCS resins under tested conditions.
Keywords:
X‐ray CTadeno‐associated virus (AAV)affinity resinchromatographycolumn foulingconfocal laser scanning microscopy (CLSM)imagingMore Related Videos
Related Concept Videos
Affinity Chromatography
2.8K
Affinity chromatography is a powerful technique extensively utilized for separating and purifying specific biomolecules from complex mixtures. It capitalizes on the highly selective binding between an analyte and its counterpart, such as antibody-antigen interactions. The counterpart is immobilized on the stationary phase, forming an affinity column. The stationary phase typically consists of solid support, such as agarose or porous glass beads, immobilizing the affinity ligand. The mobile...
2.8K
Types Of Column Chromatography
13.2K
The stability and compatibility of column material with samples are crucial for efficient purification in chromatographic techniques. Various operating parameters such as pH, temperature, or solvent affect the packing of the column material, thereby determining the purification efficiency. The choice of column material also plays an essential role in deciding the operating parameters and can be modified based on the proteins that need to be purified.
Gel Filtration Chromatography
When the...
Gel Filtration Chromatography
When the...
13.2K