Performance evaluation of a p-type co-axial HPGe detector: Long-term analysis of key parameters for gamma
Zahra Rakeb1, Sedigheh Sina2, Rasool Safari1
1Nuclear Engineering Department, School of Mechanical Engineering, Shiraz University, Shiraz, Iran.
Abstract:
Accurate determination of the parameters of each high purity germanium, HPGe detectors ensure the precision of quantitative results obtained from spectrum analysis. This study presents a comprehensive performance evaluation and long-term quality control assessment of a high-purity germanium (HPGe) gamma spectrometry system that has been operational for over 15 years. Key spectrometric measures were recorded, including energy resolution, peak shape ratios, asymmetry, peak-to-Compton ratio, relative efficiency, electronic noise, minimum detectable activity (MDA), and repeatability and reproducibility of the system. With the assistance of a calibrated source Co-60, the detector exhibited a commendable energy resolution of 2.01 keV (FWHM), along with good peak shape ratios (FWTM/FWHM = 1.73, FWFM/FWHM = 2.48) and a peak-to-Compton ratio of 52.2, indicating high resolution and zero spectral distortion. Relative detector efficiency was 33.88 % at 1332.5 keV, and electronic noise was negligible at 0.178 %. MDA for certain radionuclides was found to be in the range 0.048-0.605 Bq/kg, which showed good sensitivity even for low-activity peaks. Repeatability tests gave a coefficient of variation (%CV) less than 1 % for all peaks measured, while reproducibility was confirmed with t-test and F-test statistical tests. A Shewhart X‾-control chart revealed that at all times in all cases, measurements were within control limits, though a slight downward trend in the last few years suggests potential need for recalibration or maintenance. The findings corroborate the sustained conformity of the system to ISO/IEC 17025 quality standards and its sufficiency for precise and trustworthy gamma-ray analysis in radionuclide quantification and environmental monitoring.
More Related Videos
Related Concept Videos
Gas Chromatography: Types of Detectors-II
High-Performance Liquid Chromatography: Types of Detectors
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Atomic Emission Spectroscopy: Instrumentation