Cobalt Chrome Heads Underperform Ceramic and Ceramicized Metal Heads in the Younger Population: An American Joint
Akshay R Reddy1, Emilie N Miley2, Mahveen Jahan3
1College of Medicine, University of Florida, Gainesville, Florida.
Background:
The impact of fourth-generation ceramic, ceramicized metal, and antioxidant polyethylene has not been well characterized in younger patients undergoing total hip arthroplasty (THA). This study aimed to compare the revision rates of modern bearing couples in this population.
Methods:
American Joint Replacement Registry data was analyzed between January 2012 and June 2022 for patients aged 18 to 54 years who underwent a primary THA, had at least 1-year follow-up, and received one of six bearing types (ceramic on ceramic [CoC], ceramic on cross-linked polyethylene [CoP], ceramic on antioxidant polyethylene [CoAOP], ceramicized metal on cross-linked polyethylene [CMoP], cobalt chromium on CoP ["metal on polyethylene," MoP], and metal on antioxidant polyethylene). Outcomes were assessed through analysis of variance, chi-square, and Fisher exact tests. Cox proportional hazard models were conducted and reported as hazard ratios (HRs) and confidence intervals (CIs).
Results:
A total of 101,313 primary THA cases were identified; most patients received a CoP bearing (64.2%), and survivorship exceeded 96%. Also, CMoP had the lowest rate and a lower likelihood of all-cause revision compared to MoP (1.95%, n = 197; HR = 0.72; CIs: 0.59 to 0.87; P < 0.001). In addition to CMoP, CoP (HR = 0.73; CIs: 0.64 to 0.84; P < 0.001) and CoAOP (HR = 0.77; CIs: 0.64 to 0.94; P = 0.003) outperformed MoP. The lowest rate of revision for aseptic loosening was seen with CoC (P < 0.001), with CoAOP (HR = 0.41; CIs: 0.27 to 0.64; P < 0.001), CMoP (HR = 0.42; CIs: 0.56 to 0.70; P < 0.001), and CoP (HR = 0.55; CIs: 0.41 to 0.74; P < 0.001) also outperforming MoP.
Conclusions:
All-cause revision rates in this younger population are low with modern bearings. Head composition remains an important driver of survivorship. Ceramic or ceramicized metal on a polyethylene bearing should be considered the gold standard in this patient population.
More Related Videos
09:08Biotribological Testing and Analysis of Articular Cartilage Sliding against Metal for Implants
Published on: May 14, 2020
08:19Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
Published on: May 17, 2018
Related Concept Videos
Bone Disorders
Bone deposition is also affected by the levels of sex hormones like estrogen and testosterone that promote osteoblast activity and bone matrix synthesis. When the level of these hormones decreases due to aging, it causes a reduction in bone deposition. As a result, bone resorption by osteoclasts...
Knee Joint
A total of seven ligaments support the knee joint. The patellar ligament, which is also attached to the quadriceps femoris...
Mechanical Characteristics of Steel
The tension test is fundamental for determining tensile strength. In this test, a steel specimen is stretched using a gripping device until it breaks. The data collected during this test are used...
Bone Remodeling
Collar Bearings
Compact Bone
Compact bone, also called cortical bone, is the denser, stronger of the two types of bone tissue. It is found under the periosteum and in the diaphyses of long bones, where it provides support and protection. The microscopic structural unit of compact bone is called an osteon, or haversian system. Each osteon is composed of concentric rings of calcified...