From Bench to Bedside: Evaluating a Low-Cost Tongue Pressure Device Prototype for Clinical and Community Use
Myint Thu1, Boosana Kaboosaya1, Arporn Teeramongkonrasmee2
1Department of Oral and Maxillofacial Surgery, Faculty of Dentistry, Chulalongkorn University, Bangkok, Thailand.
Aim:
This study evaluated the test-retest reliability, concurrent validity, and usability of the Chulalongkorn University Prototype Tongue Pressure Instrument (CUPTI), a novel low-cost manometer for maximum tongue strength (MTS) measurement, in comparison to the standard Iowa Oral Performance Instrument (IOPI).
Materials And Methods:
Sixty healthy adults (20-80 years) were randomized into two groups in a cross-over design. Participants underwent MTS assessment using both devices. Reliability was assessed through intraclass correlation coefficients, while concurrent validity was evaluated through Pearson correlation and Bland-Altman analysis. Effect size (Cohen's d) was used to assess discriminative ability. Measurement time was compared using a paired t test, and multivariable regression was conducted to assess demographic associations.
Results:
Both CUPTI and IOPI demonstrated excellent test-retest reliability (ICC > 0.95). CUPTI consistently recorded lower MTS values, with a mean bias of -8.12 kPa (limit of agreement: -2.52 to 18.76 kPa). A strong correlation (r = 0.93, P < 0.001) was observed between devices. The effect size of 0.61 indicated a moderate, clinically meaningful difference. CUPTI required slightly more time than IOPI for MTS assessment (P = 0.005). Age was the only significant predictor of MTS (P = 0.034).
Conclusion:
CUPTI is a valid and reliable alternative to IOPI, offering affordability and accessibility for clinical and community dysphagia screening. Its performance supports implementation in resource-constrained settings and home-based rehabilitation.
Related Concept Videos
Assessment of the Mouth
Mouth Inspection
The inspection begins with visually examining the mouth for symmetry, color, and size.
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...
The Tongue and Taste Buds
Suctioning the Nasopharyngeal Airway
Equipment Required