Surachat Jaroonpipatkul1, Apichai Pokawattana1
1Division of Cardiology, Rajavithi Hospital, College of Medicine Rangsit University, Bangkok, Thailand.
Background:
Brugada syndrome is associated with life-threatening arrhythmias and requires implantable cardioverter-defibrillator (ICD) placement, but inappropriate shocks remain a clinical challenge.
Case Summary:
A 43-year-old male firefighter with Brugada syndrome and an automated ICD experienced both appropriate and inappropriate shocks. The shocks occurred primarily during duty in high-temperature environments. Device interrogation revealed the false shocks were due to T-wave oversensing triggered by elevated body temperature. True ventricular tachycardia episodes were also recorded during sleep. The ICD was reprogrammed, and temperature exposure precautions were advised.
Discussion:
This case demonstrates the impact of environmental heat exposure on electrocardiogram morphology and ICD performance in patients with Brugada syndrome. It emphasizes the need for individualized programming and occupational risk assessment.
Take-Home Messages:
T-wave oversensing from elevated temperature can trigger inappropriate ICD shocks in patients with Brugada syndrome. Individualized ICD settings and environmental risk management are essential in high-risk occupations.
Related Concept Videos
Dysrhythmias IV: Characteristics of Bradyarrhythmias
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Cardiomyopathy II: Dilated Cardiomyopathy
Mechanism of Cardiac Arrhythmias
Cardiomyopathy III: Hypertrophic Cardiomyopathy