Validation of performance qualification of ultrasound probe high-level disinfection devices in clinical settings
David Bellamy1, Karen Vickery2
1Sterilising Services Department (SSD) Manager, St George Hospital, World Federation for Hospital Sterilization Sciences (WFHSS), Australia.
Background:
In accordance with AS5369:2023, disinfection processes need to undergo routine performance qualification (PQ) as part of in-field validation. Automated high-level disinfection (HLD) devices therefore need to undergo PQ. Responding to reports that some ultraviolet-C (UV-C) devices used for reprocessing of ultrasound transducers were failing to achieve sporicidal efficacy during routine PQ, we evaluated the sporicidal efficacy of these devices.
Methods:
Sporicidal efficacy was assessed using commercially available stainless steel biological indicators (BIs) inoculated with 106Geobacillus stearothermophilus spores (ATCC® 7953). BIs were clamped in top and bottom positions inside the chambers of the devices [UV-C light-emitting diode (LED), UV-C lamp and hydrogen peroxide (H2O2) mist]. All three devices were in clinical use. BI test conditions included packaged, unwrapped and non-flamed as well as unwrapped and flame sterilised on the clamped coupon end. Results were evaluated on a pass (no growth) or fail (growth) basis.
Results:
The UV-C LED device failed to inactivate spores in all tested positions and conditions (n = 18). The UV-C lamp device passed 2/6 of the unwrapped and flamed tests but failed all other tests (n = 12). The H2O2 mist device passed all tests, inactivating spores for all conditions and chamber positions (n = 18).
Conclusion:
The H2O2 mist device was the only device in this study that passed all tests, demonstrating sporicidal efficacy. Both UV-C devices failed to demonstrate sporicidal efficacy in this PQ study suggesting that clinicians need to monitor UV-C devices more frequently to ensure continued efficacy and patient safety.
Related Concept Videos
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Hand hygiene
Hand washing...
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Physical Methods for Controlling Microbial Growth: Radiation and Filtration