Comparison of Agreement and Accuracy using Binocular Wavefront Optometer with Autorefractor and Phoropter
Published on: September 16, 2025
Accuracy of Reported Sizes of the EVO/EVO+ Visian Implantable Collamer® Lenses
Matthew Hirabayashi1, Gurpal Virdi2, Taj Nasser3
1Ophthalmology, Parkhurst NuVision LASIK Eye Surgery, San Antonio, USA.
Abstract:
Purpose The purpose of this study is to assess the variability between published manufacturer sizes and actual sizes measured in the operating room (OR) for Implantable Collamer® Lenses (ICLs) (STAAR Surgical, Monrovia, CA). Methods A prospective observational study was performed where the diagonal length of 50 ICLs from 50 consecutive cases was measured between June 26, 2023, and November 1, 2023. ICLs were positioned "face down" next to an OR ruler and photographed using the NGENUITY® 3D Visualization System (Alcon, Geneva, Switzerland). These images were processed in Adobe Photoshop 24.6 (Adobe Inc., San Jose, CA) to eliminate perspective distortions, and the images were scaled using known dimensions of the OR ruler. Accurate diagonal measurements were taken using the caliper tool. Measurements were repeated across ICLs in the other direction to ensure consistency. Average size discrepancy, defined as measured ICL size minus published ICL size, was reported along with standard deviation (SD) and confidence interval (CI). A two one-sided t-test (TOST) was used to compare statistical equivalence between spherical and toric ICLs within a 0.017 mm margin. Results TOST analysis revealed that the mean measured size of all lens sizes (12.1 mm, 12.6 mm, and 13.2 mm) and types (spherical and toric) is statistically equivalent (α = 0.05) to their labeled size, indicating a high level of manufacturing precision. No ICLs deviated from published measurements by more than 0.05 mm. Conclusion The actual sizes of ICLs closely match the published manufacturer sizes, suggesting high manufacturing precision. There is no significant variability between spherical and toric ICLs, supporting the reliability of published sizing for clinical decisions.
Related Concept Videos
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Angle Closure Glaucoma: Treatment
Glaucoma: Overview
Cell Size
Surface Area
Cells can take in nutrients and water via diffusion through the plasma membrane itself or through specific channels in the membrane. The area of the membrane surrounding...