Measuring shoulder abduction strength using 2 different dynamometers: comprehensive intrarater and interrater
Ecenur Atli1, Mahir Topaloglu2, Zeynep Hosbay3,4
1Department of Physiotherapy and Rehabilitation, Graduate Education Institute, Biruni University, İstanbul, Türkiye.
Acta Orthopaedica Et Traumatologica Turcica
|September 25, 2025
Summary
Handheld dynamometer (HHD) shows excellent reliability for measuring shoulder abductor isometric strength. While HHD is valid, a fixed dynamometer may be better if participants are stronger than testers.
Area of Science:
- Orthopedics
- Musculoskeletal Research
- Rehabilitation Medicine
Background:
- Accurate assessment of isometric muscle strength is crucial for diagnosing and managing shoulder conditions.
- Handheld dynamometers (HHD) offer a portable and accessible method for muscle strength testing.
- Comparing HHD with fixed dynamometers (FD) is essential to validate its clinical utility.
Purpose of the Study:
- To evaluate the intrarater and interrater reliability of HHD for shoulder abductor isometric strength.
- To compare HHD measurements with those obtained from a fixed dynamometer (FD).
Main Methods:
- Twenty-two healthy participants (age >18) were assessed twice by two experienced testers.
- Measurements were taken at 90 degrees of shoulder abduction in the scapular plane using both HHD and FD.
- Intraclass correlation coefficient (ICC), standard error of measurement, and minimal detectable change were calculated.
Main Results:
- Strong correlations were found between HHD and FD measurements for both testers (r=0.772 and r=0.748).
- Excellent intrarater reliability (ICC=0.941) and interrater reliability (ICC=0.889) were observed for HHD.
- The handheld dynamometer demonstrated high validity and reliability in assessing shoulder abductor strength.
Conclusions:
- Handheld dynamometer is a reliable and valid tool for assessing shoulder abductor muscle strength in research and clinical settings.
- Consider a fixed dynamometer if the study population's muscle strength exceeds that of the testers.
- Further research with diverse populations and tester experience levels is recommended.
Related Concept Videos
Reliability and Validity
13.7K
Reliability and validity are two important considerations that must be made with any type of data collection. Reliability refers to the ability to consistently produce a given result. In the context of psychological research, this would mean that any instruments or tools used to collect data do so in consistent, reproducible ways.
13.7K
Muscles of the Shoulder
8.9K
The muscles surrounding the shoulder girdle, including the clavicle and scapula, primarily stabilize the scapula. This stable base allows other muscles to move the humerus effectively. Scapular movements often mirror those of the humerus and extend its range of motion. For instance, raising the arm above the head would not be feasible without simultaneous upward rotation of the scapula.
Anterior Thoracic Muscles
The anterior thoracic muscles include the serratus anterior, subclavius, and...
Anterior Thoracic Muscles
The anterior thoracic muscles include the serratus anterior, subclavius, and...
8.9K