Comparison of Ocular Wavefront in Seated and Supine Positions Using a Hand-Held Hartmann-Shack Aberrometer
Noh Eun Kwon1, Nicolas Brown2, Jong Hwa Jun1
1Department of Ophthalmology, Keimyung University School of Medicine, Dongsan Medical Center, Daegu 42601, Republic of Korea.
Journal of Clinical Medicine
|September 27, 2025
Summary
Intraoperative aberrometry using a hand-held device shows consistent ocular aberration measurements whether patients are seated or supine. This supports using preoperative seated measurements for supine cataract and refractive surgeries.
Area of Science:
- Ophthalmology
- Optical Engineering
- Biomedical Optics
Background:
- Intraoperative aberrometry is crucial for cataract and corneal refractive surgery.
- Ocular aberrations are typically measured preoperatively in a seated position, but surgeries are performed supine.
- Discrepancies in measurements due to patient position may impact surgical outcomes.
Purpose of the Study:
- To compare ocular wavefront aberrations between seated and supine positions using a hand-held Hartmann-Shack aberrometer.
- To assess the repeatability and potential measurement errors of a hand-held aberrometer in different positions.
- To evaluate the feasibility of using seated measurements for supine intraoperative aberrometry.
Main Methods:
- Ocular wavefront aberrations were measured in 18 healthy subjects (36 eyes) using a hand-held Hartmann-Shack aberrometer.
- Measurements were taken five times consecutively for a 6 mm pupil under three conditions: seated (holder-fixed and hand-held) and supine (hand-held).
- Repeatability was assessed using standard deviation of repeated measurements (Srm) and intraclass correlation coefficient (ICC).
Main Results:
- No significant differences in spherical or cylindrical aberrations were found across seated and supine conditions.
- Intraclass correlation coefficient (ICC) values exceeded 0.9 for both lower-order aberrations (LOAs) and higher-order aberrations (HOAs), indicating excellent repeatability.
- Mean higher-order aberrations (HOA_RMS) showed no statistically significant difference between seated and supine positions (p=0.913).
Conclusions:
- A hand-held Hartmann-Shack aberrometer provides repeatable measurements for LOAs and HOAs.
- Supine ocular aberration measurements are comparable to seated measurements.
- Preoperative seated aberrometry results can be reliably applied to intraoperative aberrometry, irrespective of patient position.