Acceleration of slice encoding for metal artifact correction at 0.55 T using hexagonal sampling.
Bahadır Alp Barlas1, Kübra Keskin1, Bochao Li2
1Ming Hsieh Department of Electrical and Computer Engineering, University of Southern California, Los Angeles, California, USA.
Hexagonal sampling accelerates metal artifact correction (SEMAC) MRI scans by an additional 50% at 0.55T. This method reduces scan time without significant artifacts, benefiting mid-field MRI near implants.
Area of Science:
- Medical Imaging
- Magnetic Resonance Imaging (MRI)
- Biomedical Engineering
Background:
- Mid-field MRI systems (0.1T-1.0T) offer potential for imaging near metal implants.
- Limited parallel imaging options on these systems necessitate scan time reduction strategies.
- Slice Encoding for Metal Artifact Correction (SEMAC) is a key technique for metal artifact reduction.
Purpose of the Study:
- To evaluate a hexagonal sampling approach for accelerated SEMAC at 0.55T.
- To assess the feasibility of hexagonal undersampling for reducing scan times in SEMAC.
- To determine if hexagonal sampling impacts image quality or introduces artifacts.
Main Methods:
- Retrospective hexagonal undersampling was applied to SEMAC with 2-fold generalized autocalibrating partially parallel acquisitions (GPAA) at 0.55T.
- Phantom experiments with spinal fixation and hip arthroplasty hardware were conducted.
- In vivo experiments included patients with spinal fusions and a volunteer with a hip arthroplasty.
Main Results:
- Hexagonal sampling achieved an additional 50% scan time reduction.
- Image quality remained compatible with existing methods.
- Tradeoffs included minor signal void and position-dependent signal-to-noise ratio reduction at field-of-view edges.
Conclusions:
- Hexagonal sampling provides a 50% scan time reduction for SEMAC at 0.55T when combined with GPAA/parallel imaging.
- This method does not introduce substantial artifacts.
- Hexagonal sampling is a valuable tool for reducing scan times on mid-field MRI systems where parallel imaging is limited.