Cardiac Magnetic Resonance Imaging at 7 Tesla
Published on: January 6, 2019
Quantitative Confounder Analysis of Electrocardiogram Signals in Cardiac Magnetic Resonance at 1.5, 3 and
Richard Hickstein1,2, Stephanie Wiesemann1, Darian Viezzer1,2
1Working Group Cardiovascular Magnetic Resonance, Experimental and Clinical Research Center, Charité Medical Faculty, Max-Delbrück Center for Molecular Medicine, Helios Klinikum Berlin Buch, Department of Cardiology and Nephrology, Charité-Universitätsmedizin Berlin, Kardiologie-ECRC, Berlin, Germany.
Background:
The electrocardiogram (ECG) used for gating in cardiac MRI may be compromised by multiple confounders inside the scanner bore.
Purpose:
To quantify the influence of magnetic field strengths (1.5 T/3 T/7 T), standardized electrode positions, and imaging sequences on ECG signals used for gating.
Study Type:
Prospective.
Population:
Sixteen healthy volunteers (eight male; mean age 26.25 ± 7.67 years).
Field Strength/Sequence:
Balanced steady-state free precession cine (1.5 T/3 T), fast low-angle shot cine (7 T), and 4D flow (1.5 T/3 T/7 T) sequences.
Assessment:
ECG-signals were recorded during breath-hold and non-breath-hold short axis cine (sax-bh and sax-nbh, respectively) and 4D flow scans at 1.5 T/3 T/7 T. All scans were repeated with 4 standardized electrode positionings (pos1-4) at each field strength. Pos1/2 were vendor-recommended positionings for 1.5 T/3 T/7 T scans, respectively, whereas pos3/4 were alternative positionings recommended in previous studies. Similarity between confounded ECG-signals and unconfounded baseline ECG-signals was assessed by QRS-feature correlation. Cine image quality (IQ) was assessed by 3 readers (with 6, 10, and 22 years experience) on a four-point Likert scale.
Statistical Tests:
Linear mixed models with type III tests of fixed effects (overall) and t tests with adjusted degrees of freedom (pairwise subgroup-comparisons) at significance level p < 0.05.
Results:
Increasing field strength resulted in significantly decreasing similarity to baseline measurements, with r values (provided with 95% confidence interval) of 1.5 T: 97% (92.6-101.3); 3 T: 91.4% (87.1-95.8); 7 T: 50.4% (46-54.9) and lower IQ: 1.5 T: 2.33 (2.12-2.55); 3 T: 1.96 (1.75-2.17); 7 T: 0.91 (0.7-1.12). Vendor-specified electrode positions pos1: 91.8% (87.2-96.5), pos2: 88.3% (83.7-92.9) showed significantly higher correlation with baseline measurements than alternative positions pos3: 67.5% (62.9-72.1) and pos4: 70.8% (66.2-75.4). The evaluated standardized sequences showed similar amounts of electrocardiogram distortion, with r values of: sax-bh: 77.3% (73-81.7); 4D: 79.3% (75-83.7), p = 0.54; sax-nbh: 82.1% (77.8-86.5), p = 0.31, but the difference between sax-bh and sax-nbh: 4.8% (2.88-6.72) was significant.
Data Conclusion:
Increasing field strength leads to significant ECG signal distortions. Vendor-specified positions 1/2 resulted in less distorted ECG signals than alternative positions 3/4 recommended in previous publications.
Technical Efficacy:
Stage 5.
More Related Videos
07:21Magnetic Resonance Derived Myocardial Strain Assessment Using Feature Tracking
Published on: February 12, 2011
18:11A Research Method For Detecting Transient Myocardial Ischemia In Patients With Suspected Acute Coronary Syndrome Using Continuous ST-segment Analysis
Published on: December 28, 2012
Related Concept Videos
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...