Laser-based pulse oximetry eliminates pigmentation effects on oxygen saturation measurements: A pilot study
Jonas A Pologe1, Nathan K You1, Mark Blumstein1
1Research and Development, Zynex Monitoring Solutions, Boulder, Colorado, United States of America.
Background:
Almost since its introduction, pulse oximetry has been plagued by inaccuracy associated with pigmentation, whether from fingernail polishes or melanin. The presence of melanin in the optical path of a pulse oximetry sensor has been shown to artifactually increase oxygen saturation measurements which can result, clinically, in occult hypoxemia and misdiagnoses.
Methods:
This report describes the theoretical basis for this inaccuracy and presents results from both a benchtop study and a clinical study testing the effects of pigmentation on conventional LED-based pulse oximetry compared to laser-based pulse oximetry. The clinical portion of this study was performed on 18 consenting participants, nine darkly pigmented and nine lightly pigmented, to assess the ability of laser-based pulse oximetry to eliminate this dangerous pigmentation bias. The clinical study directly compared oxygen saturation readings on laser-based pulse oximeters to readings performed on two different LED-based pulse oximeters. All measurements were compared to invasive reference laboratory measurements performed on arterial blood samples. We hypothesized that monochromatic light sources used in laser-based pulse oximetry would make this new technology insensitive to pigmentation bias.
Results:
The clinical portion of this study showed significantly greater (p < 0.001) measurement error (variance) for the two LED-based pulse oximeters (5.48 and 5.47) compared to laser-based pulse oximetry (3.54), when analyzed for all participants. The bias differences in oxygen saturation measurements by the LED-based pulse oximeters, when made on lightly versus darkly pigmented participants compared to invasive reference measurements, accounts for most of the increase in measurement error.
Conclusions:
By combining theoretical development, experimental benchtop testing, and a clinical study, this research explains and demonstrates that the wide spectral bandwidth of LEDs is the root cause of pigmentation bias in commercially available LED-based pulse oximetry and validates the ability of narrow-band laser-based pulse oximetry to eliminate this pigmentation bias.
More Related Videos
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.