Objective:
To assess the environmental impact of MR and CT imaging using life cycle assessment, focusing on energy use, resource consumption, and emissions.
Methods:
This International Organization for Standardization 14040-guided life cycle assessment-based study focused on MRI and CT services, including the production and use of three MRI and four CT scanners, at a quaternary care 800-bed academic medical center in the Southeastern United States over a 1-year period. Data were collected through direct observation, record review, staff interviews, and energy metering. Environmental impacts were assessed using SimaPro 9.3.0.2 and the Ecoinvent v3.8 database.
Results:
MRI and CT services at this site generate an estimated 221 and 108 tons of carbon dioxide equivalent per year, respectively. This is equivalent to the emissions of 52 (MRI) and 25 (CT) cars driven annually. Energy consumption accounted for the largest portion of emissions (58% for MRI, 33% for CT), followed by disposable supplies (26% for MRI, 16% for CT), capital equipment production (7% for MRI, 13% for CT), and linens (4% for MRI, 11% for CT). Switching to solar photovoltaic electricity could reduce total MRI emissions by 70% and CT emissions by 40%, increasing the relative impacts of scanner production, disposable supplies, and linens.
Discussion:
This study highlights the significant environmental impact of MRI and CT services, particularly energy consumption. Renewable energy sources, such as solar photovoltaics, offer the greatest potential for mitigating the environmental footprint. Additional strategies include optimizing scanner utilization, adopting reusable or reprocessable supplies, and embracing circular business practices such as circular manufacturing and extending the lifespan of capital equipment.
More Related Videos
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System IV: CMRI
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Brain Imaging
These technologies include computerized axial tomography (CAT or CT scans), positron-emission tomography (PET scans), magnetic resonance imaging (MRI), functional magnetic resonance imaging (fMRI), and Transcranial Magnetic...
Radiological Investigation I: X-ray and CT