Mitigating banding artifacts in balanced steady-state free precession using parallel transmission in a single
Chia-Yin Wu1,2,3,4, Jin Jin1,5, Markus Barth1,2,3
1Centre for Advanced Imaging, The University of Queensland, St. Lucia, Queensland, Australia.
Purpose:
Using parallel transmission (pTx) in ultra-high field MRI allows efficient radiofrequency pulse designs that can produce uniform excitations. Here, we demonstrate that pTx can simultaneously mitigate non-uniformity and off-resonance artifacts in balanced steady-state free precession sequences at 7 T.
Theory And Methods:
The standard hard excitation pulses of the balanced steady-state free precession sequence were replaced with two pTx pulses, played out in alternating repetition times. Based on the kT-point trajectory, each pulse was designed to produce spatially varying phase distributions that compensate for off-resonance phase variations accumulated over one repetition time duration, while also mitigating non-uniformities. Depending on the local ΔB0, the different steady-state signals formed were distributed over two aliased images with a half field-of-view shift relative to one another. Using slice-GRAPPA, these images were disentangled and recombined to produce artifact-free images.
Results:
Simulations validated the concept showing that a pair of strategically designed pTx pulses can impose a favorable distribution of spatially varying steady-state. As a proof of principle, experiments performed demonstrated the technical feasibility of the concept in a phantom in the presence of induced phase variations and in vivo.
Conclusion:
With a tailored pair of pTx pulses, it was possible to concurrently mitigate and ΔB0 non-uniformities in a thin slab through the brain in a single acquisition. Although further work is needed to extend the coverage and robustness of this method before being adopted more widely, it highlights the potential of utilizing pTx capabilities beyond correction.
Related Concept Videos
Bandpass Sampling
A bandpass signal has a spectrum with a lower frequency limit, denoted as ω1, and an upper frequency limit, denoted as ω2....
Double Resonance Techniques: Overview
Spin decoupling is usually achieved by...
NMR Spectrometers: Resolution and Error Correction
¹³C NMR: ¹H–¹³C Decoupling
A broadband decoupling technique is used to simplify these complex, sometimes overlapping, signals. Broadband decoupling relies on a...