Related Experiment Video
Updated: Jun 11, 2026

Catheter Ablation in Combination With Left Atrial Appendage Closure for Atrial Fibrillation
Published on: February 26, 2013
Removal of a third-generation left atrial appendage occlusion device from the pararenal aorta using a novel
Chase Green1, Snehita Bonthu2, Timothy C Oskin2,3
1University of North Texas Health Science Center, Texas College of Osteopathic Medicine, Fort Worth, TX.
Left atrial appendage occluder devices (LAAODs) offer stroke prevention in patients with atrial fibrillation who have bleeding complications or contraindications to oral anticoagulation. We present a case of successful endovascular removal of a third-generation LAAOD embolized to the pararenal aorta. The patient presented asymptomatically after unrelated imaging demonstrated the LAAOD near the lumbar spine. The LAAOD was removed using a laparoscopic biopsy grasper via a percutaneous right femoral approach, thus avoiding a more complicated open procedure.
Left atrial appendage occluder devices (LAAODs) offer stroke prevention in patients with atrial fibrillation who have bleeding complications or contraindications to oral anticoagulation. We present a case of successful endovascular removal of a third-generation LAAOD embolized to the pararenal aorta. The patient presented asymptomatically after unrelated imaging demonstrated the LAAOD near the lumbar spine. The LAAOD was removed using a laparoscopic biopsy grasper via a percutaneous right femoral approach, thus avoiding a more complicated open procedure.
Related Concept Videos
Cardiac Catheterization III: Left Heart Catheterization
Mitral Stenosis III: Medical Management
Aneurysm III: Interprofessional Care