Application of Ultrasound and Shear Wave Elastography Imaging in a Rat Model of NAFLD/NASH
Published on: April 20, 2021
Ultrasonic radio frequency signals-based hepatorenal index improved the diagnostic performance in steatotic liver
Yunlin Huang1, Xinliang Xu1, Chao Sun2
1Department of Ultrasound, Xinhua Hospital Affiliated to Shanghai Jiao Tong University School of Medicine, Shanghai, China.
Purpose:
To evaluate the feasibility and diagnostic performance of HepatoRenal index plus (HRI+) measurements based on ultrasonic radio frequency (RF) signals, compared to auto-hepatorenal index (HRI) measurements based on B-mode ultrasound (BMUS) images.
Methods:
In this prospective study, patients referred for evaluation of suspected steatotic liver disease by abdominal ultrasound or liver biopsy were enrolled. All patients underwent HRI+ and auto-HRI measurements using a Resona A20 ultrasound system (Mindray, China). Pearson or Spearman correlation coefficient was performed based on the normality of distribution. Bland-Altman analysis and area under the curve (AUC) were performed.
Results:
From November 2023 to December 2024, 303 patients were included. The median age was 43 years (interquartile range [IQR]: 34-61), and 51.8% (157/303) of the patients were female. The median body mass index was 25.8 kg/m2 (IQR: 22.8-29.0). HRI+ values had a very low correlation with skin-to-liver capsule distance (r = 0.13) and did not correlate with the depth of regions of interest (ROIs). Auto-HRI values correlated with skin-to-liver capsule distance (r = 0.43) and depth of ROIs (r = 0.20). HRI+ and auto-HRI values had a mean bias of -0.17 and a correlation coefficient of 0.61. In a subgroup of 43 patients with histological results, HRI+ measurements showed higher diagnostic performance in detecting moderate and severe hepatic steatosis (AUCs = 0.77-0.91) compared to auto-HRI measurements (AUCs = 0.44-0.69). The cutoff values of HRI+ for diagnosing hepatic steatosis grades ≥ S1, ≥ S2, and S3 were 1.03, 1.11, and 1.22, respectively.
Conclusions:
HRI+ measurements were not affected by measurement depth and image post-processing. Compared to auto-HRI measurements, HRI+ measurements demonstrate higher feasibility and diagnostic performance in assessing hepatic steatosis.
More Related Videos
05:56Implementation of Non-invasive Point of Care Transient Elastography for Evaluation of Liver Disease in Pediatric Populations with Cystic Fibrosis
Published on: August 29, 2025
06:09Author Spotlight: Advancing Hepatic Fibrosis Diagnosis Using Magnetic Resonance Elastography and AI
Published on: July 21, 2023
Related Concept Videos
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Effect of Hepatic Disease on Pharmacokinetics: Pathophysiologic Assessment and Liver Function Test
Imaging Studies II: Ultrasonography
Ultrasonography
During an ultrasonography procedure, a handheld device called...