Performance of the MANTA Vascular Closure Device in Endovascular Aortic Procedures: A Single-Center Experience
Maksym Katelenets1, Anatoly Leytsin1, Dmitry Shepovalov1
1Department of Vascular Surgery, Soroka Medical Center, Be'er-Sheva, Israel.
Background:
Endovascular procedures requiring large-bore access demand reliable closure methods. The MANTA device offers an alternative to surgical cut-downs and suture-based techniques, with growing evidence of efficacy.
Methods:
We retrospectively analyzed 128 patients undergoing aortic endovascular repair using the MANTA device between June 2020 and November 2024. Device selection, anatomical data, perioperative technique, and outcomes were reviewed. Primary endpoints included technical success and perioperative complications; secondary endpoints assessed 30-day access-related events.
Results:
Across 228 deployments, the technical success rate was 93.9%. Perioperative complication rate was 6.6%, with major events comprising 1.8%. Bleeding and limb ischemia were the most frequent issues. False aneurysm formation occurred in 3% of cases; spontaneous thrombosis was seen in most.
Conclusion:
The MANTA device exhibits high technical success with an acceptable safety profile, supporting its use in aortic interventions and compatibility with early discharge protocols.