Related Experiment Video
Updated: May 6, 2026

Skeletal Muscle Neurovascular Coupling, Oxidative Capacity, and Microvascular Function with 'One Stop Shop' Near-infrared Spectroscopy
Published on: February 20, 2018
Single-distance, phase-only frequency-domain NIRS for vascular oxygenation and tissue metabolism: a Monte Carlo
Muaaz Faiyazuddin1, Haiyang Tang1, Yiqing Hu1
1University of Texas at Arlington, Department of Bioengineering, Arlington, Texas, United States.
Significance:
Frequency-domain near-infrared spectroscopy (FD-NIRS) currently enables absolute hemoglobin quantification but requires multidistance measurements of both amplitude attenuation and phase shifts. Notably, existing FD-NIRS approaches have not demonstrated reliable quantification of differential redox-state concentrations of cytochrome c oxidase ( ), a critical metabolic marker.
Aim:
We aimed to develop a novel optimization-based algorithm for single source-detector (S-D), phase-only FD-NIRS that achieves accurate quantifications of hemoglobin parameters (HbO and Hb) and .
Approach:
Our computational framework implemented both forward modeling and inverse reconstruction. For the modeling, we first defined chromophore concentration sets (HbO, Hb, ), followed by calculations of wavelength-dependent optical properties for two- or eight-wavelength configurations. Next, time-domain photon propagation was generated via Monte Carlo (MC) simulations, and FD-NIRS parameters (modulation amplitude, phase) were extracted through Fourier analysis. In the inverse computation, nonlinear optimization with edge-barrier regularization was employed for the recovery of chromophore concentrations. Both the multiseparation method and the single S-D, phase-only algorithm were used to reconstruct chromophore concentrations.
Results:
Respective performances evaluated for the two methods were compared through their concentration recovery accuracy. In either the two- or eight-wavelength configuration, our new algorithm outperformed the conventional method for the S-D separations up to 3 cm for all three chromophores. In particular, estimation was improved markedly from a mean relative error of 34.1% with the conventional method to just 5.1% using our algorithm.
Conclusions:
These results validate single-separation phase-only FD-NIRS as an accurate method for multichromophore quantification (including ), enabling simpler, cost-effective systems without compromising metabolic imaging capability. The approach achieves error in hemoglobin quantification while eliminating traditional multidistance requirements.