Volume targeted mask ventilation during simulated neonatal resuscitation - A randomized crossover manikin study
Chelsea Morin1,2, Kashmala Yousafzai1, Brenda Hiu Yan Law1,2
1Centre for the Studies of Asphyxia and Resuscitation, Neonatal Research Unit, Royal Alexandra Hospital, Edmonton, Alberta, Canada.
Volume-targeted ventilation (VTV-PPV) with the NextStep™ device significantly reduced mask leak in neonatal resuscitation simulations. This novel approach proved more effective than traditional T-piece positive pressure ventilation (PPV).
Area of Science:
- Neonatal Resuscitation
- Pediatric Critical Care
- Respiratory Support Devices
Background:
- Positive pressure ventilation (PPV) is crucial in neonatal resuscitation.
- Optimizing mask seal and minimizing air leak are critical for effective PPV.
- Current methods like T-piece ventilation may have limitations in achieving consistent mask seal.
Purpose of the Study:
- To compare mask positive pressure ventilation (PPV) using ventilator-based volume-targeted ventilation (VTV) with devices like the NextStep™ and Draeger ventilator.
- To evaluate the efficacy of T-piece PPV with and without a respiratory function monitor (RFM).
- To determine the primary outcome of reduced mask leak with the NextStep™ device.
Main Methods:
- A prospective, randomized, crossover, simulation study was conducted.
- Participants were trained neonatal resuscitation providers randomized to four ventilation devices: NextStep™, Draeger ventilator, T-piece PPV with RFM visible, and T-piece PPV with RFM masked.
- The primary outcome measured was the reduction in mask leak.
Main Results:
- Volume-targeted ventilation (VTV-PPV) with the NextStep™ demonstrated significantly lower mask leak (6%) compared to the Draeger ventilator (24%), T-piece with RFM (18%), and T-piece without RFM (32%).
- Delivered tidal volume was comparable across groups, but NextStep™ showed less variation.
- Peak inflation pressure was significantly lower with VTV-PPV using the NextStep™.
Conclusions:
- In a neonatal manikin model, VTV-PPV with the NextStep™ device, utilizing a two-hand hold technique, effectively reduced mask leak.
- This suggests VTV-PPV with the NextStep™ is a promising alternative to T-piece ventilation for neonatal resuscitation.
- The NextStep™ device may improve ventilation delivery by minimizing mask leak.
Related Concept Videos
Cardiopulmonary Resuscitation II: ACLS Airway Management
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...