Multiple source-to-isocenter-distance (SID) optimization for 4πrobotic radiotherapy
Jingjie Yu1, Qifan Xu1, Shusen Jing1
1Department of Radiation Oncology, University of California San Francisco, San Francisco, CA 94115, United States of America.
This study introduces a Multi-Source-to-Isocenter Distance (SID) optimization for 4π intensity-modulated radiation therapy (IMRT) using robotic arms. This novel approach significantly improves treatment plan quality, including target coverage and organ-at-risk sparing, compared to fixed SID methods.
Area of Science:
- Medical Physics
- Radiation Oncology
- Robotics in Medicine
Background:
- Conventional 4π radiotherapy often assumes a fixed source-to-isocenter distance (SID), limiting flexibility on standard C-arm linear accelerators.
- Variable SIDs are achievable with robotic radiotherapy platforms, enabling non-coplanar and non-isocentric delivery without patient repositioning.
- Larger SIDs enhance tumor coverage and field-of-view, while shorter SIDs improve modulation resolution.
Purpose of the Study:
- To develop and evaluate a Multi-SID optimization framework for robotic 4π intensity-modulated radiation therapy (IMRT).
- To simultaneously optimize beam orientation, isocenter selection, SID selection, and fluence maps.
- To assess the dosimetric benefits of Multi-SID planning compared to fixed-SID approaches in head-and-neck cancer cases.
Main Methods:
- A Multi-SID optimization framework was created, incorporating dose fidelity, fluence smoothness, and selection sparsity objectives.
- The planning system utilized a large candidate beam pool with over 1000 orientations, 10 SID levels (50-95 cm), and 5-9 isocenters.
- Multi-SID IMRT plans were compared against fixed-SID (50 cm and 100 cm) plans for 10 head-and-neck cancer patients.
Main Results:
- Multi-SID plans consistently demonstrated superior quality across all evaluated patients.
- Key dosimetric improvements included enhanced target volume homogeneity (D95/D5) and dose conformity (Paddick CI).
- Statistically significant improvements in organ-at-risk (OAR) sparing and reduced high-dose regions (R50) were observed with the Multi-SID method.
Conclusions:
- The Multi-SID optimization framework effectively leverages variable SIDs to achieve superior dosimetric outcomes in robotic 4π IMRT.
- This approach offers better target coverage and modulation resolution compared to traditional fixed-SID techniques.
- Multi-SID planning represents a significant advancement for optimizing radiotherapy treatment plans, potentially leading to more effective and efficient cancer treatment.