Related Experiment Video
Updated: Jan 14, 2026

Upper-extremity Approach for Secondary Access in Transfemoral Transcatheter Aortic Valve Implantation
Published on: August 8, 2025
30-Day Outcomes of a Controlled-Bending Delivery System for TAVR in Severe Aortic Stenosis
Yang Chen1, Zhongyu Wang2, Baihua Sun3
1Coronary Heart Disease Center, National Center for Cardiovascular Diseases and Fuwai Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China; Department of Cardiology, Peking University People's Hospital, Beijing, China.
Abstract:
This report presents the first-in-human experience with a 3-dimensional controlled-bending, fully retrievable delivery system for transfemoral transcatheter aortic valve replacement in severe aortic stenosis with anatomical challenge. In a prospective single-center series (September-December 2023) of 6 high-risk patients (mean age: 68.0 ± 3.9 years, 84% bicuspid valve, all with complex features such as horizontal heart or dilated ascending aorta), the device success rate was 100%, implantation was coaxial (implantation depth difference: 2.32 ± 1.53 mm), and postprocedural paravalvular leak was none to trivial. Hemodynamics improved substantially (gradient decreased from 75.5 ± 12.3 to 7.9 ± 2.1 mm Hg, effective orifice area increased from 0.61 ± 0.15 to 2.26 ± 0.34 cm2; P < 0.001). There were no deaths or strokes at 30 days postprocedure, and all patients reached NYHA functional class I or II. These early results suggest that controlled-bending delivery may enhance the precision and safety of transcatheter aortic valve replacement in complex anatomy, particularly with respect to the bicuspid valve.