Fracture Fixation in an Unstable Scaphoid Fracture Model: A Biomechanical Comparison of Double Screw Osteosynthesis
Carsten Surke1, Phoebe A Del Rosario2, Lachlan S Huntington2
1Aleris Hand och Fot, Stockholm, Sweden.
Purpose:
Rotational instability is an important contributor to nonunion in scaphoid waist fractures treated with a single compression screw. Double screw osteosynthesis demonstrates enhanced resistance to bending loads in unstable fracture models. This method may also provide better rotational strength, improving the construct's resistance to higher loads. This study aimed to evaluate and compare the torsional and cantilever bending strength of double screw fixation against single screw repairs combined with a strut graft and a wedge graft.
Methods:
Volar open wedge osteotomies were performed on 48 cadaveric scaphoids using customized surgical guides. Each specimen was randomly allocated to one of three treatment groups: (1) a single screw with strut graft, (2) a single screw with wedge graft, and (3) double screw fixation. Constructs underwent bending or torsion testing to 10° rotation using a mechanical testing apparatus. Stiffness, maximum torsional force, and load to 2.0 mm displacement were measured.
Results:
Double screw fixation constructs required more force to achieve 2.0 mm displacement during bending than single screw fixation constructs (mean difference: 78.3 N). The double screw group exhibited considerably greater maximum torque in the screw tightening direction than both the single screw and strut group (mean difference: 65.6 N·mm) and the single screw and wedge group (mean difference: 60.6 N·mm). Furthermore, it also demonstrated considerably higher maximum stiffness in the screw tightening direction compared to the single screw and strut group (mean difference: 4.8 N·mm/deg) and the single screw and wedge group (mean difference: 9.3 N·mm/deg).
Conclusions:
Double screw fixation for scaphoid waist fractures with volar bone loss yields greater rotational and bending stability than single screw and strut or single screw and wedge graft constructs.
Clinical Relevance:
The superior stability provided by the double screw osteosynthesis could eliminate the need for time-consuming corticocancellous bone grafting.
More Related Videos
06:38Establishment of a Segmental Femoral Critical-size Defect Model in Mice Stabilized by Plate Osteosynthesis
Published on: October 12, 2016
06:41A Minimally Invasive Model to Analyze Endochondral Fracture Healing in Mice Under Standardized Biomechanical Conditions
Published on: March 22, 2018
Related Concept Videos
Bone Remodeling
Fractures: Bone Repair
Minor fractures with no bone displacement are treated by immobilizing the fractured bone using a cast or splint. However, in the case of fractures with displaced bones, the broken bones are repositioned before immobilization to ensure successful healing without deformation and loss of function. The realignment of fractured bone ends is performed through a process called reduction. If the procedure...
Bone Remodeling and Repair