Related Experiment Video
Updated: Jan 13, 2026

Dynamic Lung Tumor Tracking for Stereotactic Ablative Body Radiation Therapy
Published on: June 7, 2015
Comparative Analysis of Dynamic Conformal Arc Therapy and Volumetric-modulated Arc Therapy in Lung Stereotactic Body
Jin-Kyu Kang1, Jae-Yong Jung2, Won Il Jang3
1Department of Radiation Oncology, Inje University Sanggye Paik Hospital, Seoul, Republic of Korea.
Background/Aim:
This study evaluated and compared dynamic conformal arc therapy (DCAT) and volumetric-modulated arc therapy (VMAT) in stereotactic body radiation therapy (SBRT) for peripheral lung tumors, focusing on dosimetric outcomes, treatment delivery efficiency, and plan robustness against multileaf collimator (MLC) positional errors.
Patients And Methods:
Thirty patients with peripheral lung tumors who underwent SBRT during the study period were included. All treatment plans were generated using the Monaco treatment planning system and delivered via a Versa HD linear accelerator. Dosimetric parameters - including target coverage, organ-at-risk (OAR) sparing, and plan complexity - were evaluated based on Radiation Therapy Oncology Group (RTOG) guidelines. The effect of MLC leaf positional errors on the dose distribution was systematically simulated and analyzed using a generalized equivalent uniform dose (gEUD).
Results:
DCAT demonstrated significantly lower D98%, D50%, Dmax, and homogeneity index, along with a higher conformity index than VMAT; however, all were within clinically acceptable ranges. The DCAT plans exhibited significantly lower plan complexity, fewer monitor units, and shorter delivery times compared to VMAT. Additionally, DCAT showed superior robustness to systematic MLC leaf opening and closing errors, with smaller dose deviations in both the planning target volume and OARs. No significant differences were observed between DCAT and VMAT in shift and random MLC errors.
Conclusion:
DCAT demonstrated clinically acceptable dosimetric results comparable to those of VMAT for lung tumor SBRT, while offering superior treatment efficiency and robustness to MLC positional errors. These findings suggested that DCAT is an effective alternative to VMAT in clinical practice.
Related Concept Videos
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...