Self-thresholding, 48:2 multiplexed readout circuit for a 3D position sensitive, 100 ps CTR TOF-PET detector
Shirin Pourashraf1, Joshua W Cates2, Derek R Innes1
1Department of Radiology, Molecular Imaging Program, School of Medicine, Stanford University, Stanford, CA, United States of America.
A new time-of-flight positron emission tomography (TOF-PET) detector uses a low-power charge-division circuit for precise 3D positioning and fast timing. This design enhances resolution and reduces system complexity, power, and cost for improved medical imaging.
Area of Science:
- Medical Imaging Physics
- Nuclear Instrumentation
- Positron Emission Tomography (PET)
Background:
- Precise three-dimensional (3D) position resolution is critical for advancing time-of-flight positron emission tomography (TOF-PET) detector performance.
- Existing TOF-PET systems often face challenges with system complexity, size, power consumption, and cost, limiting their widespread application.
- Achieving sub-2 mm depth of interaction (DOI) resolution and ~100 picosecond (ps) coincidence time resolution (CTR) is a key goal for next-generation PET scanners.
Purpose of the Study:
- To develop and validate a compact, low-power 3D position-sensitive scintillator readout circuit for TOF-PET detectors.
- To enable precise DOI resolution and maintain high CTR performance using a novel charge-division scheme.
- To reduce system complexity, footprint, power consumption, and cost compared to existing TOF-PET readout architectures.
Main Methods:
- Implemented a low-power, anger-based charge-division circuit with on-board real-time X/Y processing for 3D position-sensitive readout.
- Integrated custom low-noise, low-power front-end electronics featuring independently self-thresholding 48:2 multiplexed timing and energy/3D positioning circuits.
- Utilized 3x3x10 mm³ LGSO crystal elements coupled to silicon photomultipliers (SiPMs) for experimental validation of the detector design.
Main Results:
- Achieved continuous DOI resolution of approximately 1.6 ± 0.3 mm FWHM.
- Demonstrated an average CTR of 102.5 ± 2.7 ps FWHM across a range of SiPM bias voltages, with <5 ps FWHM variation.
- Obtained energy resolutions of 18% FWHM (TOT) and 13% FWHM (dynamic TOT) at 511 keV, with event detection down to 31 keV.
Conclusions:
- The developed 3D position-sensitive scintillator readout circuit effectively enables precise DOI resolution and maintains high CTR performance in TOF-PET detectors.
- The system achieves significant reductions (≥40%) in component count, footprint, power consumption, and cost compared to previous designs.
- This compact and efficient readout design offers a promising solution for next-generation, high-performance TOF-PET systems.