Impact of sensor configuration and melanin concentration on reflective pulse oximetry using Monte Carlo simulations
Maximilian Reiser1,2, Andreas Breidenassel3, Oliver Amft4,5
1Faculty of Electrical Engineering Landshut, HAW Landshut, Landshut, Germany. maximilian.reiser@haw-landshut.de.
Abstract:
We investigate the impact of melanin concentration CMel and photoplethysmography (PPG) sensor configuration on signal quality and estimation accuracy of oxygen saturation [Formula: see text]. We deploy Monte Carlo (MC) simulations of photon-skin interactions to estimate arterial oxygen saturations [Formula: see text] ranging from 70 to 100% and melanin concentration CMel ranging from 2.55 to 30.5%. We analysed the effects of red and infrared wavelengths (624 nm, 660 nm, 850 nm, and 940 nm), beam profiles (LED and VCSEL), as well as beam incidence angles of light sources (0°, 45°, and - 45°) on Perfusion Index (PI), reflective [Formula: see text] estimation, and signal-to-noise ratio (SNR) for source-detector distances ranging from 2 mm to 9 mm. Maximum PI was observed for in the red spectral range at 624 nm and in the infrared spectral range at 940 nm. In contrast, reflective pulse oximetry provided more accurate results at 660 nm and 850 nm, independent of wavelength combination. We conclude that the VCSEL beam profile at 0° is the optimal light source for a wide range of applications, as it offers a balance between PI and absolute [Formula: see text] estimations error.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.