Guided Endodontics: Three-Dimensional Planning and Template-Aided Preparation of Endodontic Access Cavities
Published on: May 24, 2022
Analysis of Factors Affecting the Accuracy of Static-guided Endodontics
Cheng Chen1, Diya Leng2, Tao Wu2
1Department of Oral & Maxillofacial Imaging, The Affiliated Stomatological Hospital of Nanjing Medical University, Nanjing, Jiangsu Province, China; Nanjing Stomatological Hospital, Affiliated Hospital of Medical School, Institute of Stomatology, Nanjing University, Nanjing, Jiangsu Province, China.
Introduction:
To evaluate how template type, sleeve height, revolutions per minute (RPM), tooth type, drilling distance, and drill diameter affect the accuracy of static-guided endodontics.
Methods:
In this in vitro study, 113 isolated premolars and molars collected from the oral surgery clinic and 10 three-dimensional printed resin teeth were mounted in a standard dental mold in their corresponding tooth positions. Cone-beam computed tomography and optical scans were obtained and co-registered. Templates were designed, and deviations were measured using Implant Studio (3Shape Unite, 3shape, Copenhagen, Denmark).
Results:
The angular deviation of the resin template (without metal sleeve) was higher than that of the metal template (P < .05). The apical linear deviation of the resin template (without metal sleeve) was significantly higher than that of the resin template (with metal sleeve) (P < .05). The coronal linear deviation of the metal template was higher than that of the resin template (with metal sleeve) (P < .05).The apical linear deviation of the resin template with 4-mm height sleeve was higher than that of the 6-mm and 8-mm height sleeves (P < .05), and there was no significant difference between the 6-mm and 8-mm height sleeves. Drill diameter, RPM, and tooth type (resin or isolated teeth) did not affect the accuracy of static guided endodontics.
Conclusions:
The apical linear deviation increased with drilling distance. Resin templates (without metal sleeves) are not recommended. The accuracy of static-guided endodontics was not affected by drill diameter, RPM, or tooth type (resin or isolated teeth).
More Related Videos
07:03Dynamic Navigation in Endodontics: Guided Access Cavity Preparation by Means of a Miniaturized Navigation System
Published on: May 5, 2022
05:54Real-Time Dynamic Navigation System for the Precise Quad-Zygomatic Implant Placement in a Patient with a Severely Atrophic Maxilla
Published on: October 18, 2021
Related Concept Videos
Common Leveling Mistakes and Errors
Errors in Taping