Metal biliary stent transection with over-the-scope clip removal system
Anthony M Rainho1, Tsola A Efejuku1, Rebecca M Haug1
1Division of Gastroenterology and Hepatology, University of Virginia Health, Charlottesville, Virginia, USA.
Summary
Trimming frayed uncovered self-expandable metal stents (UC-SEMSs) is challenging. A bipolar over-the-scope clip (OTSC) removal system offers a safe and effective alternative for UC-SEMS transection.
Area of Science:
- Gastroenterology
- Endoscopy
- Interventional Endoscopy
Background:
- Uncovered self-expandable metal stents (UC-SEMSs) are crucial for unresectable pancreaticobiliary malignancies.
- Fraying of UC-SEMSs can lead to stent malfunction, ulceration, and perforation, necessitating intervention.
- Current methods like Argon Plasma Coagulation (APC) for trimming stents are cumbersome, time-consuming, and carry risks of tissue injury.
Purpose of the Study:
- To evaluate the safety and efficacy of a bipolar over-the-scope clip (OTSC) removal system for transecting distorted UC-SEMSs.
- To present a novel, off-label approach for managing complications associated with UC-SEMSs.
Main Methods:
- Two cases of distorted UC-SEMSs requiring transection were treated using a bipolar OTSC removal system.
- The system's electrical pulses were activated upon adequate metal contact with the stent.
- Compatibility and charging requirements of the paired generator were noted.
Main Results:
- The bipolar OTSC removal system successfully transected both distorted UC-SEMSs.
- No adverse events or tissue injury were observed in either case.
- The procedure allowed for subsequent necessary interventions.
Conclusions:
- The bipolar OTSC removal system provides a safe, effective, and potentially less cumbersome alternative to APC for transecting UC-SEMSs.
- This off-label use of the OTSC removal system demonstrates its versatility in managing complex stent-related issues.
- Further studies may be warranted to standardize this technique and explore its broader applications.