Ex-Vivo Prostate Evaluation of Fused-Data TREIT Using Only Biopsy-Probe Electrodes
IEEE Transactions on Bio-Medical Engineering
|November 20, 2025
Summary
Transrectal electrical impedance tomography (TREIT) significantly improves prostate cancer imaging by increasing sensed tissue volume compared to standard biopsies. This fused-data approach shows promise for enhanced detection and characterization of prostate tumors.
Area of Science:
- Biomedical Engineering
- Medical Imaging
- Oncology
Background:
- Prostate cancer diagnosis relies on biopsies, which sample limited tissue volume.
- Electrical Impedance Tomography (EIT) offers a non-invasive imaging modality.
- Transrectal EIT (TREIT) aims to improve prostate cancer detection.
Purpose of the Study:
- To evaluate a fused-data TREIT method for enhanced prostate cancer imaging.
- To increase the sensed tissue volume compared to standard 12-core biopsies.
- To assess TREIT's utility in conjunction with pathology information.
Main Methods:
- Utilized a previously optimized TREIT algorithm on 22 ex vivo prostates.
- Implemented two TREIT approaches: EIT-P (with prostate boundary) and EIT-P+T (with tumor boundary).
- Validated TREIT against biopsy core, 3D image, and tumor-grade data.
Main Results:
- The EIT-P approach, sensing 15% of prostate tissue, significantly outperformed standard biopsies (<1%).
- Achieved best Area Under the Curve (AUC) values of 0.85, 0.84, and 0.83 for increasing tissue volumes.
- EIT-P+T showed potential for tumor-grade analysis, though further data is needed.
Conclusions:
- Fused-data TREIT substantially increases the sensed volume for prostate cancer imaging.
- TREIT represents a significant advancement over traditional biopsy and prior EIT studies.
- Further in vivo investigation of TREIT for prostate cancer is warranted.