Defining window settings for abdominopelvic photon-counting CT virtual monoenergetic imaging: quantitative and
Liqiang Ren1, Richard W Ahn1, Vasantha Vasan1
1Department of Radiology, UT Southwestern Medical Center, Dallas, TX 75390, USA.
European Journal of Radiology
|November 22, 2025
Summary
Optimizing window settings for virtual monoenergetic images (VMIs) at 60 keV on photon-counting CT (PCCT) significantly improves soft tissue and liver visualization. This enhancement is crucial for accurate abdominopelvic imaging with reduced radiation dose.
Area of Science:
- Medical Imaging
- Radiology
- Photon-Counting CT Technology
Background:
- Virtual monoenergetic images (VMIs) at 60 keV are recommended for photon-counting CT (PCCT) soft tissue imaging, offering reduced radiation dose.
- Default windowing settings on 60 keV VMIs result in images appearing brighter than standard 120-kV CT, necessitating manual adjustments for optimal visualization.
Purpose of the Study:
- To determine optimal soft tissue and liver window settings for abdominopelvic imaging using PCCT at 60 and 70 keV.
- To quantitatively and qualitatively evaluate the impact of default versus optimized window settings on image quality.
Main Methods:
- Retrospective analysis of dual-phase abdominopelvic PCCT 60 and 70 keV VMIs from 39 patients.
- Radiologists manually adjusted window center/width settings for optimal soft tissue and liver visualization.
- Quantification of contrast-to-noise ratio (CNR) and display-CNR (dCNR) across different window settings; reader evaluation of image preference and overall impression.
Main Results:
- Mean CNR was significantly higher at 60 keV compared to 70 keV (p < 0.001).
- Optimized windowing at 60 keV substantially increased liver dCNR (75-92%) and soft tissue dCNR (27-51%) compared to default settings.
- Images viewed with optimal windowing at 60 keV received significantly higher reader rankings and impressions for both soft tissue and liver evaluation (p < 0.001).
Conclusions:
- Optimized window settings are critical for enhancing soft tissue and liver visualization on PCCT VMIs.
- The benefits of optimized windowing are particularly pronounced at 60 keV, improving diagnostic quality in abdominopelvic imaging.