Evaluation of the small-field output factor in eclipse modeling methods using representative beam and measured data
Kunio Nishioka1, Yuki Kunii1, Yoshinori Tanabe2
1Department of Radiology, Tokuyama Central Hospital, 1-1 Kodacho, Shunan, Yamaguchi, 745-8522, Japan.
None:
Beam modeling for radiotherapy treatment planning systems (RTPS) can be performed using representative beam data (RBD) or direct measurements. However, RBD typically excludes output factor (OPF) measurements for fields smaller than 3 × 3 cm2. The Eclipse treatment planning system addresses this limitation by incorporating measured OPF data for fields as small as 1 × 1 cm2. Although existing studies have primarily examined the accuracy of small-field OPFs for plastic scintillator detectors, studies directly comparing the OPF values obtained through RBD modeling with and without OPF measurements for small field sizes are limited. Therefore, this study proposes a novel measurement approach using data averaged from an ion chamber and diode detector for small-field dosimetry to provide critical insights into the integration of OPFs for these small field sizes in RBD-based beam modeling. We systematically evaluated the impact of small-field OPF measurements on beam modeling accuracy by comparing three distinct approaches: (1) RBD-based modeling without small-field OPF data, (2) RBD-based modeling incorporating measured small-field OPF data, and (3) modeling based solely on measured data, with and without the inclusion of 1 × 1 cm2 field sizes. In addition, we compared OPF values obtained from a W2 plastic scintillator detector with the averaged OPF values from a PinPoint 3D ion chamber and EDGE diode detector across multiple beam energies and flattening filter-free (FFF) configurations. Our analysis included field sizes ranging from 1 × 1 cm2 to 40 × 40 cm2. The results demonstrated that for square fields, OPF calculation differences between RBD modeling with and without measured data were < 1.5%, < 4.5%, and < 4.5% at 1 × 1 cm2, and < 0.5%, < 1.5%, and < 1.5% at 2 × 2 cm2, respectively. The RBD group exhibited a trend in which the OPF difference increased with the expansion of the irradiation field size. Notably, the most significant variations between modeling approaches occurred along the upper jaw expansion direction in rectangular fields. This suggests that a thorough evaluation is necessary for modeling results with an OPF ≤ 1 × 1 cm2. This study highlights the advantages and disadvantages of beam modeling using measured OPF and RBD, providing valuable insights for future facilities that rely solely on RBD for beam modeling.
More Related Videos
07:31Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
06:58Electrochemical Etching and Characterization of Sharp Field Emission Points for Electron Impact Ionization
Published on: July 12, 2016
Related Concept Videos
Atomic Emission Spectroscopy: Lab
Atomic Emission Spectroscopy: Instrumentation
Atomic Emission Spectroscopy: Overview
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....