Related Experiment Video
Updated: Jan 10, 2026

Dynamic Lung Tumor Tracking for Stereotactic Ablative Body Radiation Therapy
Published on: June 7, 2015
Marker-Less Lung Tumor Tracking from Real-Time Color X-Ray Fluoroscopic Images Using Cross-Patient Deep Learning
Yongxuan Yan1, Fumitake Fujii1, Takehiro Shiinoki2
1Graduate School of Sciences and Technology for Innovation, Yamaguchi University, Ube 755-8611, Japan.
Abstract:
Fiducial marker implantation for tumor localization in radiotherapy is effective but invasive and carries complication risks. To address this, we propose a marker-less tumor tracking framework to explore the feasibility of a cross-patient deep learning model, aiming to eliminate the need for per-patient retraining. A novel degradation model generates realistic simulated data from digitally reconstructed radiographs (DRRs) to train a Restormer network, which transforms clinical fluoroscopic images into clean, DRR-like images. Subsequently, a DUCK-Net model, trained on DRRs, performs tumor segmentation. We conducted a feasibility study using a clinical dataset from 7 lung cancer patients, comprising 100 distinct treatment fields. The framework achieved an average processing time of 179.8 ms per image and demonstrated high accuracy: the median 3D Euclidean tumor center tracking error was 1.53 mm, with directional errors of 0.98±0.70 mm (LR), 1.09±0.74 mm (SI), and 1.34±0.94 mm (AP). These promising results validate our approach as a proof-of-concept for a cross-patient marker-less tumor tracking solution, though further large-scale validation is required to confirm broad clinical applicability.