Comparison of Distortion-Product Otoacoustic Emissions Measured in the Same Subjects Using Four Commercial Systems
Edyta Pilka1,2, Henryk Skarżyński1,2, W Wiktor Jedrzejczak1,2
1Institute of Physiology and Pathology of Hearing, Mochnackiego 10 Street, 02-042 Warsaw, Poland.
None:
Background/Objectives: Distortion-product otoacoustic emissions (DPOAEs) are suited to longitudinal cochlear assessment, but inter-system differences may confound interpretation across clinics. This study compared DPOAE outcomes across four commercial systems and evaluated their within-session repeatability. Methods: Adults with normal hearing (84 ears) were tested using the HearID DP (Mimosa Acoustics), SmartDPOAE (Intelligent Hearing Systems), Eclipse DPOAE20 (Interacoustics), and Echoport ILO 292 USB I (Otodynamics). DPOAEs were recorded at 1, 1.5, 2, 3, 4, 6 and 8 kHz using a criterion of ≥6 dB signal-to-noise ratio. Two measurements per ear were obtained, with the probe repositioned between sessions. Results: All systems showed similar frequency response profiles but substantially different absolute values. Between-system amplitude differences were smallest at 1.5-4 kHz and largest at 6 kHz. Noise floors varied considerably: HearID DP and SmartDPOAE were best (lowest) while Echoport ILO 292 USB I and Eclipse DPOAE20 were worst (highest), with inter-system differences most prominent between 1.5 and 4 kHz. HearID DP achieved the highest detection rates (84/84 ears at key frequencies). Test-retest reliability was good across all systems. The standard error of measurement varied from 0.99 to 2.88 dB, the smallest being the HearID DP. Within-session differences were typically ≤2 dB, with the best repeatability between 1.5 and 6 kHz. Conclusions: Despite similar frequency responses, clinically significant differences exist between DPOAE systems in terms of noise floors, signal-to-noise ratios, and response amplitudes. Inter-device variations frequently exceeded minimum detectable change values, meaning that DPOAE devices cannot be considered clinically interchangeable. These findings underscore the need for industry-wide standards to enable reliable cross-clinic comparisons.
Related Concept Videos
Echo
Imagine the sound is reflected back to the ears. Assuming that the source is very close to the human, the difference between hearing the two sounds—the emitted sound and the reflected sound—may be more than the minimum time for perceiving distinct sounds. If this is the case,...
Electronic Distance Measuring Instruments