Fourier Transform Near-Infrared Spectrometry Reveals Significant Inter-Lot Variability in Thyrotropin Alfa
Uiyeol Yoon1,2, Ayumi Tarianto3, Philip J Almeter1,2
1Department of Pharmacy Services, University of Kentucky HealthCare, Lexington, KY 40536.
Summary
Thyrotropin alfa, a recombinant human thyroid stimulating hormone (TSH), exhibits significant spectral variability within and between manufacturing lots. This variability, detected via near-infrared spectroscopy, may indicate potential differences in drug stability and efficacy.
Area of Science:
- Biochemistry
- Analytical Chemistry
- Pharmaceutical Science
Background:
- Thyrotropin alfa is a recombinant glycoprotein and a purified form of human thyroid stimulating hormone (TSH).
- It is produced using recombinant DNA technology in genetically modified Chinese hamster ovary (CHO) cells.
- The drug is supplied as a lyophilized product for intramuscular administration.
Purpose of the Study:
- To assess the spectral consistency of thyrotropin alfa within and between manufacturing lots.
- To identify potential variations in drug product quality using near-infrared (NIR) spectroscopy.
- To investigate if spectral differences correlate with drug stability and efficacy.
Main Methods:
- Intralot and interlot analysis of thyrotropin alfa vials using near-infrared (NIR) spectroscopy.
- Multivariate analysis of spectral data to identify outliers and variability.
- Comparison of spectral plots across different wavenumber regions.
Main Results:
- Significant spectral differences were observed within a single manufacturing lot (EY0155), with 25% of vials flagged as outliers.
- Interlot analysis of 171 vials from 18 lots revealed considerable variability in NIR absorption profiles.
- Specific wavenumber regions (e.g., 4000-5000 cm⁻¹, 5150 cm⁻¹, 5900 cm⁻¹, 6650 cm⁻¹) showed notable differences, potentially related to moisture content and peak ratios.
Conclusions:
- Substantial spectral variability exists in thyrotropin alfa products, both within and between manufacturing lots.
- These observed spectral differences may have implications for the drug's stability and therapeutic efficacy.
- Further investigation is warranted to understand the clinical significance of these variations.
Related Concept Videos
IR Frequency Region: Fingerprint Region
1.8K
IR spectra are divided into two main regions: the diagnostic region and the fingerprint region. The diagnostic region of the spectrum lies above 1500 cm−1. The absorptions resulting from single-bond vibrations of the N–H, C–H, and O–H stretch at higher wavenumbers and appear on the left side of the spectrum. The stretching absorptions of the C≡C and C≡N occur between 2100–2300 cm−1. In contrast, those arising from stretching absorptions of the...
1.8K
IR Spectrometers
2.2K
There are two main infrared (IR) spectrophotometers: dispersive IR spectrometers and Fourier transform infrared (FTIR) spectrometers. In a dispersive IR spectrometer, a beam of infrared radiation produced by a hot wire is divided into two parallel equal-intensity beams using mirrors. One beam passes through the sample, while another is a reference beam. The beams then move through the monochromator, which separates the radiations into a continuous spectrum of different frequencies. The...