Comparison of spectral performance of three dual-energy CT scanners equipped with a deep-learning image
Joël Greffier1, Cécile Salvat2, Maxime Pastor1
1IMAGINE UR UM 103, Montpellier University, Department of Medical Imaging, Nîmes University Hospital, 30900 Nîmes, France.
Purpose:
The purpose of this study was to compare the spectral performance of three dual-energy CT (DECT) scanners and one photon-counting CT (PCCT) scanner on virtual monoenergetic images (VMIs) at low-energy levels and on iodine maps.
Materials And Methods:
A spectral phantom was scanned using one PCCT scanner and three different DECT scanners that included a rapid kV-switching CT (R-KVSCT), an ultrafast kV-switching (U-KVSCT) and a dual-layer CT (DLCT) scanner. Acquisitions were obtained with each CT system using classical abdominal and pelvic examination parameters, as well as a volume CT dose index at 11 mGy. VMI at 40/50/60/70 keV and iodine maps were reconstructed for each scanner. Noise power spectrum (NPS) and task-based transfer function (TTF) were evaluated. Detectability indexes (d') were computed to model the detection task of two contrast-enhanced lesions.
Results:
Noise magnitude decreased from 40 to 70 keV for all DECT scanners and this decrease was greater for R-KVSCT (-80.0 ± 0.1 [standard deviation (SD)] %) and less pronounced for DLCT (-14.4 ± 0.8 [SD] %) scanners. The average NPS spatial frequency (fav) values decreased from 40 to 70 keV (0.26 to 0.17 mm-1) for R-KVSCT, increased for DLCT (0.18 to 0.25 mm-1) but were similar for U-KVSCT (0.19 ± 0.002 [SD] mm-1) and PCCT (0.21 ± [SD] 0.008 mm-1) scanner. For R-KVSCT and PCCT scanners, TTF at 50 % (f50) values increased from 40 to 70 keV for both inserts. For U-KVSCT and DLCT scanners, similar f50 values were found according to energy level for both inserts. For both contrast-enhanced lesions, d' values decreased from 40 to 70 keV for PCCT, DLCT and U-KVSCT scanners. For R-KVSCT scanner, d' values peaked at 60 keV. At 40 and 50 keV, the greatest d' values were found with DLCT and PCCT scanners.
Conclusion:
At 40 or 50 keV, the best combined results (objective and subjective assessments) are obtained with DLCT and PCCT scanners.