Comparative efficacy of decontamination methods for laparoscopic equipment: a systematic review and meta-analysis
Jiajing Li1,2,3, Shibin Gao4, SuBo Zhang1,2,3
1Department of Medical Imaging, The Second People's Hospital of Lianyungang, Lianyungang, China.
Objective:
To compare the effectiveness of manual cleaning, alkaline multi-enzyme immersion with ultrasonic cleaning, and automatic reprocessing machines in decontaminating laparoscopes through a systematic review and meta-analysis of randomised controlled trials.
Methods:
A comprehensive literature search was conducted across PubMed, Embase, Cochrane Library, Web of Science, Sinomed, CNKI and Wanfang databases from inception to February 2025. Randomised controlled trials comparing different cleaning and disinfection methods for laparoscopes were included. The primary outcome was the qualified rate of decontamination, defined as meeting predetermined thresholds for each detection method (visual cleanliness, protein <6.4 μg/cm², ATP <200 RLU, negative occult blood). Heterogeneity was assessed using I² statistics, with subgroup analyses by detection method and intervention type. Risk of bias was evaluated using the Cochrane risk of bias tool.
Results:
Eleven randomised controlled trials involving 4,661 cases were included. Meta-analysis showed that alkaline multi-enzyme immersion with ultrasonic cleaning improved qualified decontamination rates compared with manual cleaning alone when assessed by visual inspection (risk ratio [RR] = 1.07, 95% CI: 1.02-1.13, P < 0.01) and occult blood test (RR = 1.12, 95% CI: 1.02-1.23, P < 0.05). The 7% improvement in first-pass cleaning qualification translates to potentially preventing contamination in approximately 70 additional instruments per 1,000 processed. Automatic reprocessing machines showed similar improvements (RR = 1.08, 95% CI: 1.01-1.16, P < 0.05). Low heterogeneity (I² < 25%) was observed across most outcomes.
Conclusion:
The evidence suggests that combined cleaning methods provide modest but clinically meaningful improvements over manual cleaning alone, though certainty is limited by methodological constraints and geographic concentration of studies. Healthcare facilities should consider implementing enhanced protocols while weighing resource availability, training requirements, and local infection prevention priorities.
Related Concept Videos
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...